If any problems are found during the initial setup procedure, refer to the 

troubleshooting section on page 13. For further assistance, contact 

Drucker Diagnostics at 814-692-7661. 

1.

 

Plug the centrifuge in to an approved electrical outlet. For 

electrical safety, the unit must always be properly grounded. 

2.

 

For safety purposes, the locking system is always activated. To 

deactivate the system to insert or retrieve samples, press the 

‘OPEN / STOP’ button on the control panel. The ‘UNLOCKED’ 

indicator light should illuminate. If it does not, refer to 

Troubleshooting section. The lid will be unlocked for 15 seconds 

after pushing the ‘OPEN / STOP’ button. 

3.

 

Turn the latch counterclockwise and open the lid. 

4.

 

Spin the rotor by hand; check for free and level rotation. If the 

rotor does not spin freely, refer to Troubleshooting section. 

5.

 

Place the six test tube holders inside the rotor and verify that they 

are seated properly. 

6.

 

Close the lid. Rotate the lid knob clockwise to its complete top 

position. The ’LATCHED’ indicator light should be illuminated. If it 

is not, make sure that the lid is latched properly. The centrifuge 

will not run unless the lid is latched and that the ’LATCHED’ light is 

on. 

7.

 

Turn the centrifuge on by pushing the ‘START’ button. 

8.

 

The ‘RUNNING’ indicator light will illuminate. 

9.

 

The test tube holders will slide up into the Fixed Angle position 

and the unit will accelerate to full speed. 

10.

 

Listen to the sound of the centrifuge. A smooth whirring sound 

should be heard. If there are any loud or unusual sounds, stop the 

centrifuge by pushing the ’OPEN / STOP’ button immediately and 

refer to Troubleshooting section. 

11.

 

While the machine is running, try to turn the latch 

counterclockwise. Power may be cut to the motor, but you should 

be unable to fully turn the latch. If it is possible to turn the latch 

and open the lid while the unit is running, contact Drucker 

Diagnostics for assistance. Close and latch the lid. 

12.

 

Push the ‘OPEN / STOP’ button. The ‘RUNNING’ indicator light 

should go out and the motor should slow to a stop. 

Summary of Contents for 614VET

Page 1: ...614VET Operator s Manual P N 03 0 0002 0229 Rev C...

Page 2: ...5 Initial Setup Procedure 6 Control Panel 7 Verifying The Preset Time 8 Changing The Preset Time 8 Operation 9 Spinning Balanced Loads 10 Rotor Removal And Installation 10 To remove the rotor 10 To i...

Page 3: ...e holders inside machine MODEL DESCRIPTION The Model 614VET is a continuous duty electronically controlled Fixed Angle centrifuge with a lid safety interlock system The unit is controlled by an electr...

Page 4: ...flow design that prevents overheating of samples Heavy gauge steel construction for safety and durability Lid safety switch that prevents the centrifuge from operating unless the lid is closed and la...

Page 5: ...uipment is present 2 Choose a setup location which meets the following criteria a A clearance height of 20 is required to open the lid b The clearance envelope is the space around the centrifuge which...

Page 6: ...the rotor and verify that they are seated properly 6 Close the lid Rotate the lid knob clockwise to its complete top position The LATCHED indicator light should be illuminated If it is not make sure...

Page 7: ...for fifteen 15 additional seconds After the centrifuge has passed this procedure it is ready for operation CONTROL PANEL RUNNING Lights up when the machine is in operation power is being applied to t...

Page 8: ...CHANGING THE PRESET TIME NOTE Your centrifuge must be plugged in a Push the OPEN STOP button to disengage the lock and then open the lid b Push and hold the START and OPEN buttons for approximately th...

Page 9: ...of two 2 minutes To display or change this time setting refer to page 8 6 Turn on the machine by pushing the START button on the control panel 7 The centrifuge should begin to spin The RUNNING indicat...

Page 10: ...plug the centrifuge from the electrical outlet to eliminate the possibility of electrical shock or other injury 2 Remove the test tube holders 3 Remove the nut in the center of the rotor by turning it...

Page 11: ...e safety guidelines of your laboratory to properly clean up and or dispose of materials in the event that a substance known to be potentially toxic radioactive or contaminated with a pathogenic microo...

Page 12: ...in water or other cleaning solutions as this will cause damage and void the warranty 3 ONLY isopropyl alcohol soap and water or a 10 5500 PPM bleach solution should be used for cleaning and disinfect...

Page 13: ...replace it if necessary The centrifuge does not run o Check the electrical outlet o Make sure the lid latch is turned completely clockwise to its stop position When the lid is closed properly the lat...

Page 14: ...t to the desired length Check the run preset by following the instructions on page 8 If the preset is not the desired length follow the procedure on the same page to change the run preset time EMERGEN...

Page 15: ...14VET has a true 0 RPM lid locking system The lid safety locking system keeps the lid locked at all times even during power failure and requires that the rotor be at rest in order to unlock the lid Th...

Page 16: ...d 7751043 Circuit Breaker 7760002 Power cord 03 1 0008 0016 Pawl latch lid 03 1 0008 0009 Knob latch lid 7724071 Hinge friction 7732018 Seal lid gasket 7713032 125mm Tube Holder Black 02 002 1 0027 Li...

Reviews: