Drucker Diagnostics 614VET Operator'S Manual Download Page 10

Drucker Diagn1-814-692-7661 - [email protected]

 

Page | 10

  

S

PINNING 

B

ALANCED 

L

OADS

 

Your centrifuge must contain a balanced load in order to work properly. 

Use the following rules when loading the rotor. 

Spinning balanced loads will extend the life of the machine and produce 

better results. 

 

1.

 

Opposing tube holders must be identical and must contain the 

same cushion, or none at all. 

2.

 

Opposing tube holders must be empty or loaded with equally 

weighted samples. 

3.

 

If an odd number of samples is to be spun, fill a tube with water 

to match the weight of the unpaired sample and place it across 

from this sample. 

R

OTOR 

R

EMOVAL 

A

ND 

I

NSTALLATION

 

T

O REMOVE THE ROTOR

1.

 

Unlock the centrifuge by pushing the ‘OPEN / STOP’ button and 

unlatch and open the lid. 

CAUTION: 

Unplug the centrifuge from the electrical outlet to 

eliminate the possibility of electrical shock or other injury. 

2.

 

Remove the test tube holders. 

3.

 

Remove the nut in the center of the rotor by turning it 

counterclockwise (a tool may be required). 

4.

 

The rotor is sitting on a cone-shaped adapter. Pull the rotor up 

and off of this adapter. 

T

O INSTALL THE ROTOR

1.

 

Place the rotor back onto the cone-shaped adapter. You may 

need to turn the rotor slightly to line it up properly. 

2.

 

The rotor should slide onto the rotor cone freely. 

3.

 

Once a proper fit has been achieved, replace the nut and turn it 

until it is hand-tight, (a tool may be required). 

4.

 

Replace the tube holders and verify that they are seated 

properly. 

5.

 

It is recommended that the initial setup procedures be 

performed to ensure that the rotor has been installed correctly 

and that no damage has been done to the centrifuge during 

either the rotor installation or possible rotor chamber cleaning. 

See page 6 for this procedure. 

Summary of Contents for 614VET

Page 1: ...614VET Operator s Manual P N 03 0 0002 0229 Rev C...

Page 2: ...5 Initial Setup Procedure 6 Control Panel 7 Verifying The Preset Time 8 Changing The Preset Time 8 Operation 9 Spinning Balanced Loads 10 Rotor Removal And Installation 10 To remove the rotor 10 To i...

Page 3: ...e holders inside machine MODEL DESCRIPTION The Model 614VET is a continuous duty electronically controlled Fixed Angle centrifuge with a lid safety interlock system The unit is controlled by an electr...

Page 4: ...flow design that prevents overheating of samples Heavy gauge steel construction for safety and durability Lid safety switch that prevents the centrifuge from operating unless the lid is closed and la...

Page 5: ...uipment is present 2 Choose a setup location which meets the following criteria a A clearance height of 20 is required to open the lid b The clearance envelope is the space around the centrifuge which...

Page 6: ...the rotor and verify that they are seated properly 6 Close the lid Rotate the lid knob clockwise to its complete top position The LATCHED indicator light should be illuminated If it is not make sure...

Page 7: ...for fifteen 15 additional seconds After the centrifuge has passed this procedure it is ready for operation CONTROL PANEL RUNNING Lights up when the machine is in operation power is being applied to t...

Page 8: ...CHANGING THE PRESET TIME NOTE Your centrifuge must be plugged in a Push the OPEN STOP button to disengage the lock and then open the lid b Push and hold the START and OPEN buttons for approximately th...

Page 9: ...of two 2 minutes To display or change this time setting refer to page 8 6 Turn on the machine by pushing the START button on the control panel 7 The centrifuge should begin to spin The RUNNING indicat...

Page 10: ...plug the centrifuge from the electrical outlet to eliminate the possibility of electrical shock or other injury 2 Remove the test tube holders 3 Remove the nut in the center of the rotor by turning it...

Page 11: ...e safety guidelines of your laboratory to properly clean up and or dispose of materials in the event that a substance known to be potentially toxic radioactive or contaminated with a pathogenic microo...

Page 12: ...in water or other cleaning solutions as this will cause damage and void the warranty 3 ONLY isopropyl alcohol soap and water or a 10 5500 PPM bleach solution should be used for cleaning and disinfect...

Page 13: ...replace it if necessary The centrifuge does not run o Check the electrical outlet o Make sure the lid latch is turned completely clockwise to its stop position When the lid is closed properly the lat...

Page 14: ...t to the desired length Check the run preset by following the instructions on page 8 If the preset is not the desired length follow the procedure on the same page to change the run preset time EMERGEN...

Page 15: ...14VET has a true 0 RPM lid locking system The lid safety locking system keeps the lid locked at all times even during power failure and requires that the rotor be at rest in order to unlock the lid Th...

Page 16: ...d 7751043 Circuit Breaker 7760002 Power cord 03 1 0008 0016 Pawl latch lid 03 1 0008 0009 Knob latch lid 7724071 Hinge friction 7732018 Seal lid gasket 7713032 125mm Tube Holder Black 02 002 1 0027 Li...

Reviews: