Oximeter Series
PULSE OXIMETER INSTRUCTION
MANUAL
32
Alarm limits range: 70%-100%
Default limits: High 99%; low 90%
SpO
2
Display range: 0%~100%
Measurement range: 70%~100%
Resolution: 1%
Accuracy: ±3% (70-100%) Unspecified (0-69%)
Measurement Wavelengths and Output Power
Red 660nm @ 3mw nominal
Infrared 940nm @3mw nominal
Pulse Rate
Display range: 0~254bpm
Measurement range: 30~235bpm
Resolution: 1bpm
Accuracy: 30~99bpm: ±2bpm; 100~235bpm: ±2%
Operation Environment
Operating temperature: 5
℃
~40
℃
Relative humidity: ≤ RH80%, no condensation
Atmosphere pressure: 86kPa
~
106kPa
Power supply: Two AA alkaline batteries;
Working time: work for 30 hours continuously
Abbreviations
CISPR International Special Committee on Radio Interference
EEC European Economic Community
EMC Electromagnetic Compatibility