Oximeter Series
PULSE OXIMETER INSTRUCTION
MANUAL
29
a)
.
Connect SpO
2
Probe to the SpO
2
connector on the oximeter.
b)
.
Insert the operator’s finger into the finger sensor, the SpO
2
measured value of healthy person should be from 95% to 100%,
and the pulse rate is same as heart rate,
c)
.
If SpO
2
Simulator is available, verify the accuracy of Oxygen
Saturation Value with probes as follows:
Oxygen Saturation Tolerance
96% ±2%
86% ±2%
70% ±3%
9.2
.4 SpO
2
& Pulse Rate Alarm Verification
a). Connect SpO
2
Probe to the SpO
2
connector on the oximeter.
b)
. Insert the operator’s finger into the finger sensor, the SpO
2
measured value of healthy person should be more than 95%.
c). Set the SpO
2
high limit as 90, low limit as 80.
d). Verify the SpO
2
visual and auditory alarms, the background color
of the SpO
2
data should be red and beep sound should be heard.
9.3 Trouble Shooting
a) Can’t power on the oximeter
Please check the batteries voltage.