![Dressler CESAR 1312 User Manual Download Page 124](http://html1.mh-extra.com/html/dressler/cesar-1312/cesar-1312_user-manual_2530911124.webp)
5-4
Installation, Setup, and Operation
CESAR 1312 (E)
Dressler
®
HF-Technik GmbH
Grounding
Connect the ground terminal at the rear panel of the CESAR generator with
the nearest PE ground (usually the 19” rack). Use a proper grounding cable
that meets your national standards and has a cross-sectional area of at least
4 mm
2
. Proper operation requires that you use a central, common ground for
the generator, matching network, and load.
Connecting RF Output Power
The following illustration shows the output power connector, and
describes each pin.
Figure 5-1. Output (RF Out) connector
To Connect RF Output:
1. Connect one end of a shielded coaxial cable of 50
Ω
impedance to the
RF
Out
connector.
Do not attempt to turn on power until the CESAR generator is grounded.
Table 5-1. Output connector pin descriptions
Pin
Description
Center
RF output
Outer
Ground
Summary of Contents for CESAR 1312
Page 1: ...U s e r M a n u a l s e r v i c e d r e s s l e r c o m CESAR Generator Model 1312 ...
Page 2: ......
Page 3: ...User Manual CESAR Generator Model 1312 CESAR 1312 E ...
Page 6: ...iv CESAR 1312 E Dressler HF Technik GmbH ...
Page 14: ...xii List of Tables CESAR 1312 E Dressler HF Technik GmbH ...
Page 24: ...1 10 Safety and Product Compliance Guidelines CESAR 1312 E Dressler HF Technik GmbH ...
Page 34: ...3 8 Specifications CESAR 1312 E Dressler HF Technik GmbH ...
Page 170: ...5 50 Installation Setup and Operation CESAR 1312 E Dressler HF Technik GmbH ...
Page 188: ...6 18 Troubleshooting and Customer Support CESAR 1312 E Dressler HF Technik GmbH ...