VIII
Gentier Real-Time PCR System
Reminding:
under normal circumstances, please use the instrument attached power cord.
If the original power cord is destroyed, please substitute it with an equal one.
Caution
: for protection against electric shock hazards, the Gentier instrument must be
grounded properly. The power cord provided is a standard three-pin plug, please plug it
into an appropriate three-wire grounded receptacle for operation safety.
Caution
: the power grid environment of Gentier instrument must have ground wire.
Caution
: please check the power connection carefully. Hold the power plug when you plug
the power cord and make sure the power plug is perfectly inserted into the socket, do not
pull the power cord to pull out the plug.
Caution
: please keep the power cord away from heater or other high temperature objects.
Please do not put anything on the power cord and keep it away from places where people
move around.
Caution
: the fuse tube type of the Gentier instrument is 250V~/F10AH, located in the fuse
tube box near the power outlet at the rear of the instrument. Use improper fuse tube may
lead to circuit system damage. Please check and make sure that the fuse tube has been
properly installed before switch on the instrument.
Caution
: before replacing the fuse tube, please cut off the instrument power supply, unplug
the power plug and use a screwdriver to pry open the fuse box. Then substitute the old fuse
tube with an equal one.
IV. Environmental Safeties and Precautions
Caution
: Gentier instrument is for indoor use only, the room should be well ventilated and
without corrosive gas.
Prohibit
: never run the Gentier instrument in places that have or may have flammable and
explosive gas.
Prohibit
: spraying liquid on electrical parts may cause a short circuit and result in fire, do
not use sprays in vicinity of the Gentier instrument.
Reminding
: the working environment temperature of Gentier instrument should be
between 10°C~30°C, and the relative humidity should be between 20%~85%.