52
Gentier Real-Time PCR System
Figure D-6.4 Sample Setting interface - Auto Concentration Setting Window
User can set the starting concentration for a series of diluted standard samples in the
Starting Conc
. input box;
User can select the concentration dilution factor for a series of diluted standard samples
from
Dilution Facto
r drop-down list;
User can select the dye for the corresponding standard curve of a series of diluted standard
samples from
Dye
drop-down list;
User can select the number of standard samples from
Replicate
drop-down list;
User can select the auto concentration setting direction from
Direction
drop-down list:
Horizontal
(from left to right) and
Vertical
(from up to down);
User can click
< OK >
and the application software will conform to the auto concentration
setting direction, calculate and display the concentration of all the selected standard
samples.
User can click
< Cancel >
and the application software will give up the current auto
concentration setting and return back to the sample setting interface.