90
Gentier Real-Time PCR System
threshold for the X axis and Y axis of scatter plot.
c.
Quality Threshold
: user can use
and
keys to set the quality threshold of
application software algorithm. Increase the quality threshold value can improve the
credibility of grouping result.
d.
Genotype Name
: user can change the grouping name of allelic gene and heterozygote gene.
II.
Gene and Sample
tab: the introduction of relevant parameters and function keys is same
with those introduced in
D-3.3.4.1Abs Quant - Abs QuantAnalysis Setting.
3.3.4.6 End Point Fluorescence
The end point fluorescence analysis displays the final detection result based on the fluorescence
intensities measured at the end point of amplification plateau phase.
The end point fluorescence analysis interface consists of seven functional modules:
Amplification Curve, End Point Fluorescence Scatter Plot, Sample Setting, Result Table, Raw
Curve, Raw Fluorescence
and
Heat Map
. The end point fluorescence analysis interface is
divided into four areas, the application software displays four functional modules
(
Amplification Curve, End Point Fluorescence Scatter Plot, Sample Setting, Result Table
) by
default, as shown in figure D-18.