31
Chapter 6 Common Fault Diagnosis Analysis and Elimination
Failure Phenomenon
Cause Analysis
Elimination
Method
Turn on the power
switch, the indicator
light is not bright or
the filament of the
tube is not bright.
Power box 2A, 5A fuse blown
Replace the fuse
The cooling water tank pump is not turned on
Turn on the water
pump and check if
the water pressure
is sufficient
The argon cylinder valve is not opened or the
argon gas is exhausted
Open the argon
valve or change the
cylinder
When the high
voltage is off, the off
indicator does not
light up
Damaged indicator
Replace the "high
voltage off" button
(the light is inside
the button)
The high-voltage electromagnetic contactor is
damaged, or the auxiliary contact wiring is
disconnected/falling off, etc.
Replace the
contactor, or
restore the normal
connection after
finding the wiring
fault
High voltage cannot
be turned on
Unplug the
high-voltage line on
the high-voltage
board of the power
box to see if the
high-voltage switch
can be turned on
No
High voltage
board failure
Contact the
manufacturer to
replace the
high-voltage board
The power
resistance in the
high frequency
box is damaged
Replace the power
resistor after
confirmation
(contact the
manufacturer for
accessories)
Damaged tube
Contact the
manufacturer to
replace the tube
Yes
High voltage filter
capacitor is
damaged
Contact the
manufacturer to
replace the filter
capacitor
Damaged choke
coil or
high-voltage
silicon stack
Contact the
manufacturer to
replace damaged
parts
The capacitor in the high-frequency box is
Contact the
Summary of Contents for DW-TY-9900
Page 16: ...15 In the new method enter the method name The following window appears determined...
Page 17: ...16 OK displays the following window...
Page 19: ...18 attenuation measurement standards measure the sample Selection Method...
Page 21: ...20 Automatic peak search...
Page 24: ...23...
Page 25: ...24 After you finish testing the standard point of exit to save Test sample...