30
the joint should be replaced when there are impurities such as metal oxides at the water pipe
joint.
5.3 Precautions for Daily Maintenance
1
.
To replace the parts installed on the working coil( such as the torch), the high-voltage power
supply must be turned off(or the ICP power supply).
2. When performing maintenance on the power box, the operator must shut it down for 30
minutes before you can handle it, so as to avoid the high-voltage electric energy stored in the
high-voltage filter capacitor from discharging, causing equipment or personal injury.
Summary of Contents for DW-TY-9900
Page 16: ...15 In the new method enter the method name The following window appears determined...
Page 17: ...16 OK displays the following window...
Page 19: ...18 attenuation measurement standards measure the sample Selection Method...
Page 21: ...20 Automatic peak search...
Page 24: ...23...
Page 25: ...24 After you finish testing the standard point of exit to save Test sample...