6-4
Transient Limit, continued
Procedure to
record transient
disturbance
PowerVisa uses three different algorithms to detect and record transients:
• Peak or Instantaneous trigger magnitude
• Waveshape threshold duration or window
• Waveshape threshold magnitude or tolerance
Continued on next page
Action...
Result...
STEP 1: From the Advanced
Options screen, press Transient
Limit.
• Press the
limit field
to change
limit. Use the keypad to enter
the instantaneous limit. This
value is compared against the
absolute value of each A/D
sample of the V and I channel
waveforms.
NOTE: Do not set this value to
less than or equal to the peak of
the normal waveform. For a
sinusoidal waveform, the peak
is 1.414 x the rms value.
Failure to do this will result in
continuous triggering.
•
Enable
toggles whether this
limit is enabled or disabled.
•
Amps
toggles between Volts
and Amps.
• Press
Group A B C
to set
identical limit value for
Channels A, B and C.
• Press
Next
to set the
waveshape threshold duration.
Proceed to Step 2 on page 6-5.
• Press
Cancel
to discard
changes and return to
Advanced Options menu.
MARK112_101
MARK112_102
Summary of Contents for PowerVisa
Page 1: ...PowerVisaTM USER S GUIDE DRANETZ BMI 1000 New Durham Road Edison New Jersey 08818 4019...
Page 13: ...xiii Dranetz BMI PowerVisaTM...
Page 14: ......
Page 78: ...4 16...
Page 136: ...6 22...
Page 158: ...7 22...
Page 204: ...A 10...
Page 221: ...C 5 APPENDIX C Battery Specifications and Replacement Procedure Battery removal diagram...
Page 222: ...C 6...
Page 224: ...D 2...
Page 248: ...F 4...
Page 259: ...MARK201 LEVEL 200 Scope Mode MARK202 MARK205 G 9...
Page 269: ...MARK112b LEVEL 112_701 Advanced Options EN50160 Options MARK112c G 19...