CONFIGURATION (continued)
Auto Plot ECG
By selecting the ON option of the Auto Plot
ECG function, the operator can choose to
have the VITALERT 2000 plot the ECG
waveform on the strip chart recorder when
one or more of the following occur:
•
Heart rate alarm
•
Preselected time interval (every 2.5, 5,
10, 15, or 30 minutes)
On power-up, the VITALERT 2000 defaults
the Auto Plot function to OFF; all criteria
selections are retained in memory.
Communication Port Configuration
This configuration screen (Fig. 9) allows the
operator to configure the serial port for
communication with the NARKOMED 3,
NARKOMED 2B, VITALERT 1000, or other
device
capable
of
communications
via
Vitalink.
The following serial port configurations are
possible:
BAUD RATE:
1200, 2400, 4800 or 9600
DATA BITS:
7 or 8
PARITY:
ODD, EVEN or NONE
STOP BITS:
1 or 2
RTS/CTS:
ON or OFF
HI
LO
P1
0
P2
SCALE
HELP
ALARMS
OFF
BLOOD PRESURE ALARMS
HEART RATE ALARMS
HI
LO
P1
0
SCALE
P2
FREEZE
OP27009
LEAD
SCALE
BASELINE AUTOSCALE
CONFIG
68
100
70
0
SELECT
ENTER
MONITOR
2400
9600
7
EVEN
1
2
ODD
NONE
VITALINK
ON
OFF
8
BAUD RATE:
DATA BITS:
PARITY:
STOP BITS:
RTS/CTS:
PROTOCOL:
CONFIGURE COMMUNICATION PORT
4800
1200
FIGURE
FIGURE 9:
9: COMMUNICATION PORT CONFIGURATION SCREEN
14