CONFIGURATION (continued)
Auto Print and Auto Plot Configuration
This Configuration screen (Fig. 8) enables
the Auto Print and Auto Plot ECG functions,
using the strip chart recorder. Auto Print
allows the VITALERT 2000 to automatically
produce a printout of the Monitor system
data values when the selected criteria are
met. Auto Plot ECG allows the VITALERT
2000 to automatically produce a plot of the
ECG waveform when the selected criteria
are met.
Auto Print
By selecting the ON option of the Auto Print
function, the operator can choose to have the
VITALERT 2000 produce a printout of
current data values on the strip chart
recorder when one or more of the following
occur:
• Warning alarm
• Caution alarm
• Noninvasive blood pressure
measurement
• Preselected time interval (every 2.5, 5, 10,
15, or 30 minutes)
On power-up, the VITALERT 2000 defaults
the Auto Print function to OFF; all criteria
selections are retained in memory.
HI
LO
P1
0
P2
SCALE
ALARMS
OFF
BLOOD PRESURE ALARMS
HEART RATE ALARMS
HI
LO
P1
0
SCALE
P2
FREEZE
OP27008
LEAD
SCALE
BASELINE AUTOSCALE
CONFIG
HELP
68
100
70
0
SELECT
MONITOR
ENTER
OFF
ON
ON
30
30
O N
ON
AUTO PRINT:
EACH WARNING ALARM:
EACH CAUTION ALARM:
EACH NIBP SAMPLE:
TIME INTERVAL:
AUTO PLOT ECG:
EACH HEART RATE ALARM:
EACH TIME INTERVAL:
CONFIGURE AUTO PRINT AND AUTO PLOT
ON
ON
OFF
OFF
OFF
OFF
OFF
15
15
10
10
5
5
30
30
2.5
2.5
OFF
OFF
FIGURE
FIGURE 8:
8: AUTO PRINT AND AUTO PLOT CONFIGURATION SCREEN
13