![Dräger VITALERT 2000 Operator'S Instruction Manual Download Page 15](http://html.mh-extra.com/html/drager/vitalert-2000/vitalert-2000_operators-instruction-manual_2526846015.webp)
INSTALLATION (continued)
4. Configure the VITALERT 2000 serial
port,
via
the
Communication
Port
Configuration screen for the following
parameters:
BAUD RATE:
2400
DATA BITS:
8
PARITY:
NONE
STOP BITS:
1
RTS/CTS:
OFF
PROTOCOL:
VITALINK
5. Configure Port A of the VITALERT 1000
for the same parameters given in step 4
(see Configuration section in
VITALERT
1000 Operators Manual
).
Connecting the VITALERT 2000 to a
NARKOMED 3
To interface the VITALERT 2000 to the
NARKOMED 3 anesthesia system (with the
five-port serial interface):
1. Connect one end of the DB9/DB9 30-inch
data cable (Part # 4110328) to the serial
port on the rear of the VITALERT 2000
(Fig. 2).
2. Connect the other end of the cable to
Port E on the rear of the NARKOMED 3
(see
NARKOMED 3 Operators Manual
).
3. Secure all connections using the captive
screws provided.
4. Configure the VITALERT 2000 serial
port,
via
the
Communication
Port
Configuration screen, for the following
parameters:
BAUD RATE:
9600
DATA BITS:
8
PARITY:
EVEN
STOP BITS:
2
RTS/CTS:
OFF
PROTOCOL:
VITALINK
Port E of the NARKOMED 3 automatically
defaults to the configuration values outlined
in Step 4.
NOTE:
In order to ensure the proper
operation of Vitalink communications, the
following firmware versions (or later) should
be present on the Narkomed 3: Trend 1.04,
Serial IF (ECC) 1.02, CCC 1.07.
Connecting the VITALERT 2000 to a
NARKOMED 2B
To interface the VITALERT 2000 to the
NARKOMED 2B anesthesia system:
1. Connect one end of the DB9/DB9 30-inch
data cable (Part # 4110328) to the serial
port on the rear of the VITALERT 2000
(Fig. 2).
2. Connect the other end of the cable to
Port A on the rear of the NARKOMED
2B (see
NARKOMED 2B Operators
Manual
).
3. Secure all connections using the captive
screws provided.
4. Configure the VITALERT 2000 serial
port,
via
the
Communication
Port
Configuration screen, for the following
parameters:
BAUD RATE:
1200
DATA BITS:
8
PARITY:
NONE
STOP BITS:
1
RTS/CTS:
OFF
PROTOCOL:
VITALINK
5. Configure Port A of the NARKOMED 2B
for the same parameters given in step 4
( s e e
C o n f i g u r a t i o n
s e c t i o n
i n
NARKOMED 2B Operators Manual
).