INSTALLATION (continued)
Connecting
the
VITALERT
2000
to
a
NARKOMED 4
To interface the VITALERT 2000 to the
NARKOMED 4 anesthesia system:
1. Connect one end of the DB9/DB9 30-inch
data cable (Part # 4110328) to the serial
port on the rear of the VITALERT 2000
(Fig. 2).
2. Connect the other end of the cable to
either Port C or Port D on the rear of
the NARKOMED 4 (see
NARKOMED 4
Operator’s Manual
).
3. Secure all connections using the captive
screws provided.
4. Configure the VITALERT 2000 serial
port,
via
the
Communication
Port
Configuration screen, for the following
parameters:
BAUD RATE:
9600
DATA BITS:
8
PARITY:
NONE
STOP BITS:
1
RTS/CTS:
ON
PROTOCOL:
VITALINK
5. Configure
Port
C
or
D
of
the
NARKOMED 4 for the same parameters
given in step 4 (see Configuration section
in
NARKOMED 4 Operator’s Manual
).
NOTE:
There is no RTS/CTS setting
with the NARKOMED 4.
Connecting
the
VITALERT
2000
to
a
NARKOMED 2C
To interface the VITALERT 2000 to the
NARKOMED 2C anesthesia system:
1. Connect one end of the DB9/DB9 30-inch
data cable (Part # 4110328) to the serial
port on the rear of the VITALERT 2000
(Fig. 2).
2. Connect the other end of the cable to
either Port A or Port B on the rear of the
NARKOMED 2C (see
NARKOMED 2C
Operator’s Manual
).
3. Secure all connections using the captive
screws provided.
4. Configure the VITALERT 2000 serial
port,
via
the
Communication
Port
Configuration screen, for the following
parameters:
BAUD RATE:
9600
DATA BITS:
8
PARITY:
NONE
STOP BITS:
1
RTS/CTS:
ON
PROTOCOL:
VITALINK
5. Configure
Port
A
or
B
of
the
NARKOMED
2C
for
the
same
parameters
given
in
step
4
(see
Configuration section in
NARKOMED
2C Operator’s Manual
).
NOTE:
There is no RTS/CTS setting
with the NARKOMED 2C.