Trouble Shooting
Messages in the alarm window
Alarm
Cause
Remedy
!!! MV low
The minute volume MV has dropped
Check patient's condition, check ventilation
below its lower alarm limit.
pattern, adjust alarm limits if necessary.
Leak in exhalation system.
Ensure connections in exhalation
system are tight.
Faulty flow sensor.
Replace flow sensor.
The ventilator may not function properly.
Contact your local DrägerService
for additional support.
!! No int. battery ?
Internal battery not installed, faulty
Fit battery or confirm alarm or change
or wrong battery installed.
internal battery.
! No int. battery ?
Internal battery not installed, faulty
Advisory message, is displayed continuously
or wrong battery installed.
when confirmed. Change internal battery.
! No int. battery
Internal battery cannot be charged.
Press the
Alarm Reset
key to confirm
charging
the alarm. Change internal battery.