Trouble Shooting
Messages in the alarm window
Alarm
Cause
Remedy
!! High respiratory rate
Patient breathes at a high
Check patient's condition, check ventilation
spontaneous rate.
pattern, correct alarm limit
RRsp
if necessary.
!! Int. battery charging inop
Technical defect.
Contact your local DrägerService for
additional support – restricted operation is
now possible.
!!! Int. battery discharged
The operating time for operation with the
The ventilator must immediately be reconnected
internal battery has expired and an
to a mains supply, an on-board DC supply
external DC supply has not been connected.
or a fully charged battery.
!! Int. battery in use
Oxylog 2000
plus
draws its power from
Press the
Alarm Reset
key to
the internal battery due to the absence of
confirm the alarm.
an external DC supply.
!! Key failed
Technical defect.
Contact your local DrägerService for
additional support – restricted operation is
now possible.