Preparation
Determining the approximate pneumatic operating time for the Oxylog® 2000
plus
Example for supply of medical gas:
• Cylinder pressure measured on the pressure gauge of the pressure reducer: 20,000 kPa (200 bar)
• Liquid capacity of the O
2
cylinder: 2.1 L
Supply of medical gas:
2.1 L x 20,000 kPa = approximately 420 L
Example for pneumatic operation time:
• VC-CMV mode, respiratory rate 10 breaths /min, VT = 1 L, O
2
= 100%
• Minute volume = 10 breaths /min x 1 L = 10L/min
Operation time =
Medical gas supply [L]
(MV +0.5*) [L/min]
Operation time = 420 = approx. 40 minutes
10.5
* Calculated with average gas consumption of ventilator: 0.5 L/min