Starting operation
Oxyl
og 2000
plus
A
Switch ON
• Briefly press the
key (A).
The Oxylog 2000
plus
performs a self-test.
• The self-test will be completed in approximately six seconds.
Upon completion of the self-test, the ventilator automatically begins
ventilation with the default settings. The opening display with configured
settings is displayed if the central rotary knob is not pressed.
The manufacturer's default settings are:
• Ventilation mode
VC-CMV
.
• Ventilation time ratio
I:E
= 1:1.5.
• Positive end expiratory pressure
PEEP
= 5 mbar.
• Plateau time
Tplat
% = 0%.
•
Trigger
= OFF.
Operation / Settings
The Oxylog 2000
plus
Pocket Guide is not a replacement or substitute for the Instructions for Use.
Any use of the device requires full understanding and strict observation of the Instructions for Use.