Alarms and Measured Values
Setting alarm limits
Oxylo
g 2000
plus
A
B
To set alarm limits for MV and RRsp
• Press the key
Alarms
(A).
Display example Alarms screen with variable alarm limits:
= lower alarm limit.
= upper alarm limit.
Set the maximum airway pressure Pmax via the Pmax control (B).
The airway pressure is limited when Pmax is reached; inspiration will
not be terminated prematurely.
Setting alarm limits automatically (Autoset)
The function Auto alarm limits sets the alarm limits on the basis of the
following actual measured values at the time of activation:
MV
: Measured value MV + 2 L/min
MV
: Measured value MV – 2 L/min
RRsp
: Measured value RRsp + 5 /min, with a minimum of 10/min.
Alarm
Range
MV
2 to 41 L/min
MV
0.5 to 40 L/min
RRsp
10 to 100 /min
This automatic selection of alarm limits is performed only once, when confirmed, via the rotary knob.