18
Technical manual
|
X-am
®
2800
Troubleshooting
3xx
EC1
6xx
EC4
4xx
EC2
Code
Cause
Remedy
150
Ambient pressure too high.
Operate gas detector within per
-
mitted ambient pressure range.
152
Customer’s life time counter will
expire soon.
Reset life time counter with Dräger
CC-Vision.
156
Ambient pressure too low.
Operate gas detector within per
-
mitted ambient pressure range.
159
Calibration not possible. Menu
function cannot be performed as a
notice exists that prevents the
function (e.g. sensors warming
up).
Determine notice code using the
Info menu and disable if neces
-
sary.
163
Life time expired.
Reset life time counter with Dräger
CC-Vision.
165
Error sent from Bluetooth control. Switch the gas detector off and
then on again / Contact DrägeSer
-
vice.
175
Bluetooth connection interrupted. Reconnect the device.
182
Device temperature too high.
Operate gas detector within per
-
mitted temperature range.
183
Device temperature too low.
Operate gas detector within per
-
mitted temperature range.
185
Battery pre-alarm
Charge or replace battery.
x51
Sensor warming up (warm-up 1).
Followed by measurement
standby, bump test standby
Wait until warm-up time is com
-
plete.
x52
Sensor warming up (warm-up 2).
Followed by calibration standby
Wait until warm-up time is com
-
plete.
x53
Value negative.
Perform a fresh air calibration.
x54
Temperature too high.
Operate gas detector within per
-
mitted temperature range.
x56
Calibration interval has expired.
Carry out the span calibration.
x69
Bump test expired.
Carry out bump test.
x70
Bump test interval expires soon.
Carry out bump test.
x71
Calibration expires soon
(CatEx thermal conduction range).
Carry out the span calibration.
x73
Calibration expired
(CatEx thermal conduction range).
Carry out the span calibration.
x77
Temperature too low.
Operate gas detector within per
-
mitted temperature range.