Supplement to the instructions for use
|
Isolette 8000 plus
7
Reprocessing
2.5 Before
reprocessing
Observe before disassembly
1. Switch off the device and all devices connected to it.
2. Disconnect all power plugs.
2.5.1
Patient-specific accessories and consumables
The patient-specific accessories and consumables must be removed from the
device and, if necessary, disassembled.
Disposable products:
●
Dispose of the disposable products.
2.5.2 Device-specific
components
The device-specific components must be removed from the device and, if
necessary, disassembled.
Removing the scale and the bed and related components
1. Disconnect the cables from the sensor module.
2. Slowly raise the hood (A).
Classification
Device-specific components
Non-critical
Device surface including controller display
Trolley and holders
Hood and walls
Gaskets, grommets
Humidity system
Removable parts of the main system
Semi-critical
None
Critical
None
49640
B
C
D
E
E
F
A