Supplement to the instructions for use
|
Isolette 8000 plus
17
Reprocessing
Prerequisites:
– The washer-disinfector has been prepared in accordance with the
manufacturer's instructions.
Positioning the components in the load carrier
1. Position the components horizontally, in an open basket or directly on the tray to
be stable.
2. Ensure the following:
– All surfaces and interiors can be completely rinsed.
– The water can drain off freely.
Performing reprocessing
1. Select a cycle.
2. Use demineralized water for the final rinse.
3. When the cycle has ended, check the components for visible soiling and repeat
the cycle if necessary.
4. Check the components for visible damage and replace if necessary.
5. Allow the product to dry completely.
2.6.4
Storage and transport
After reprocessing, there are no special requirements for storage and transport of
the product. However, the following must be observed:
– Store dry and free of dust
– Avoid recontamination and damage during transport
All further information on storage and transport included in the accompanying
documents must be observed.
2.7
Other agents and reprocessing procedures
2.7.1 Disinfectants
Use nationally approved disinfectants suitable for the respective reprocessing
process and the intended application.
Step
Agent
Manufac-
turer
Concentra-
tion
Temperature
Contact time
Cleaning
Neodisher Medi-
Clean forte
Dr. Weigert
_
Min. 90 °C
(194 °F)
Min. 5 min
1)
1) Machine run times may vary depending upon your washer-disinfector temperature settings
Flushing
Demineralized
water
–
–
Tap water tem-
perature
Min. 1 min
Disinfecting
Neodisher Medi-
Clean forte
Dr. Weigert
–
Min. 90 °C
(194 °F)
Min. 5 min
Drying
–
–
–
–
Drying time
depends on the
load