Notes on troubleshooting

Dräger DrugCheck 3000

7

8

Notes on troubleshooting

9

Storage

The Dräger DrugCheck 3000 must be stored at temperatures between

+4 °C and +25 °C (+39.2 °F and +77 °F). 

10

Disposal

The Dräger DrugCheck 3000 test kits can be disposed of in

domestic waste.

Possible fault

Possible fault cause

Preventative measures

No control line appears. 

No or low colouration of the indicator ring. Not shaken for long 

enough or strongly enough.

Follow the Instructions for Use.

Sample does not completely flow up 

the test strip. 

Dräger DrugCheck 3000 is not positioned on an even surface. 

Dräger DrugCheck 3000 has not been started when in a 

vertical position. 

The test strip has no contact with the sample liquid.

Follow the Instructions for Use.

Observe the waiting time.

Lightly tap the bottom of the test kit 

repeatedly on a flat surface.

Unclear results or blurred lines

The test strips have been soaked with too much of the sample. 

Dräger DrugCheck 3000 has not been correctly started.

Improper positioning of the Dräger DrugCheck 3000 

Dräger DrugCheck 3000 is not positioned on an even surface.

Follow the Instructions for Use. 

Colour blindness 

(for analyte result interpretation)

The result and control lines are coloured. The intensity of the 

colour is not significant as only the existence or absence of a 

line is relevant for the interpretation of the result. 

Follow the Instructions for Use.

Questionable results

Physical degeneration of the Dräger DrugCheck 3000, 

incorrect storage, opening the packaging too early before 

starting the test or reading off the results too late.

Follow the Instructions for Use for the 

correct device storage, handling and 

interpretation of the results. 

Multiple occurrence of invalid results, sample is contaminated.

Observe waiting time of 10 minutes 

before starting the sampling process.

Formation of test and control lines is 

delayed.

Temperature too low.

Indicator ring insufficiently discoloured.

Warm the Dräger DrugCheck 3000 in 

your hands before use.

Wait for the runtime of 10 minutes.

Shake the test kit until the indicator 

ring is clearly discoloured.

False-positive results at low 

temperatures.

The test lines are evaluated too early.

Wait for the runtime of 10 minutes.

Summary of Contents for DrugCheck 3000

Page 1: ...nderstand and strictly observe the instructions Use the product only for the purposes specified in the Intended use section of this document i enUS Instructions for Use 4 fr Notice d utilisation 8 es...

Page 2: ...Additional charges from your mobile service provider might occur EN How to use video on your smartphone at any time Simply scan the code...

Page 3: ...3 Dr ger DrugCheck 3000 1 2 3 6 7 5 4 C MET THC BZO C AMP OPI COC C MET THC C AMP OPI COC 1 A B 2 3 4 5 6 00233736 eps 6 2 6 3 THC 10sec 60sec 6 1...

Page 4: ...heck 3000 require professional assessment with reference to further clinical assessment of the test subjects This applies in particular to a positive not negative test result see also table 1 T1 on pa...

Page 5: ...test result Once test lines appear in the drug drug types area in the inspection window and the control lines appear as well the negative results can be read off for the corresponding drug drug type...

Page 6: ...peat the test with a new Dr ger DrugCheck 3000 test kit 7 Performance features 7 1 Influence of food and beverages Oral fluid may be contaminated immediately after consuming food beverages and or swee...

Page 7: ...ning of the Dr ger DrugCheck 3000 Dr ger DrugCheck 3000 is not positioned on an even surface Follow the Instructions for Use Colour blindness for analyte result interpretation The result and control l...

Page 8: ...gazeuse ou liquide spectrom trie de masse GC MS ou LC MS Pour tous les r sultats du Dr ger DrugCheck 3000 une valuation professionnelle des r sultats obtenus prenant en compte les informations cliniqu...

Page 9: ...ur la bandelette r active 4 Analyse 4 1 Conditions de l analyse Veiller ce que l clairage soit suffisant lors de la lecture des r sultats par ex lumi re de jour ou lampe de poche 4 2 R sultats du test...

Page 10: ...atoire alternative telle que la GC MS ou LC MS doit tre utilis e pour obtenir un r sultat analytique valid Pour tous les r sultats du Dr ger DrugCheck 3000 une valuation professionnelle des r sultats...

Page 11: ...ctement Dr ger DrugCheck 3000 mal positionn Le Dr ger DrugCheck 3000 n a pas t pos plat sur une surface plane Observer la notice d utilisation Absence de couleur pour l interpr tation du r sultat de l...

Page 12: ...DrugCheck 3000 se precisa de una evaluaci n profesional de los resultados obtenidos teniendo en cuenta la informaci n cl nica restante de la persona analizada Esto es aplicable especialmente en caso...

Page 13: ...anscurso de un minuto golpear de nuevo el kit de pruebas apoyado sobre una superficie plana Vigilar continuamente la formaci n de la l nea en la tira de an lisis 4 Evaluaci n 4 1 Condiciones para la e...

Page 14: ...poder obtener un resultado anal tico ratificado debe aplicarse un procedimiento alternativo en laboratorio m s exacto como GC MS o LC MS En todos los resultados del Dr ger DrugCheck 3000 se precisa d...

Page 15: ...n err nea del Dr ger DrugCheck 3000 El Dr ger DrugCheck 3000 no est apoyado por completo sobre una superficie lisa Observar las instrucciones de uso Daltonismo para la interpretaci n de los resultados...

Page 16: ...spectrometria de massa GC MS ou LC MS Em todos os resultados do DrugCheck 3000 da Dr ger necess rio obter um parecer profissional dos resultados fornecidos tendo em conta uma posterior avalia o cl nic...

Page 17: ...ra de teste bater o kit de teste com a superf cie de apoio numa base plana Observar continuamente a forma o de linhas na tira de teste 4 Avalia o 4 1 Condi es para a avalia o Para ler os resultados ga...

Page 18: ...o comprovado deve ser utilizado um m todo de laborat rio alternativo como a GC MS ou LC MS Em todos os resultados do DrugCheck 3000 da Dr ger necess rio obter um parecer profissional dos resultados fo...

Page 19: ...orrectamente Coloca o errada do Dr ger DrugCheck 3000 O Dr ger DrugCheck 3000 n o est bem apoiado sobre uma superf cie plana Respeitar as instru es de utiliza o Daltonismo para a interpreta o dos resu...

Page 20: ...0 JWH 18 100000 Amphetamine related compounds ng mL Dopamine 40000 MBDB 50000 MDA 40 MDEA 10000 Phentermine 100000 S Amphetamine 25 Tyramine 5000 Methamphetamine related compounds ng mL Chloroquine su...

Page 21: ...Dr ger DrugCheck 3000 21...

Page 22: ...x 49 451 882 20 80 www draeger com 90 33 737 GA 4755 210_enus_es_fr_ptbr Dr ger Safety AG Co KGaA Edition 07 September 2018 Subject to alteration Directive 98 79 EC Dr ger Safety AG Co KGaA Revalstra...

Reviews: