22
Instructions for use DigiFlow
English
English
– The software version, e.g., 2.0 (large figures),
and the date of manufacture as Year.Week, e.g.,
11.03 (small figures), are shown.
– The serial number is shown.
Operation
After the device has been prepared as described in
section "Preparation", page 21, the display shows
the following information:
The thumb wheel (B) closed completely.
The current measured value is shown in L/min*. The
required flow can be set.
The rotating flow indicator (G) means that a gas flow
is measured.
The bar display (H) shows the measured value in
relation to the measurement range of the device.
See "Technical data", page 28.
The numerical value (F) shows the flow in L/min.
Measured value
The numerical value (F) is displayed in L/min
(20 °C/1013 hPa), i.e., the numerical value refers to
an ambient temperature of 20 °C and an ambient
pressure of 1013 hPa.
To convert this value to the flow at ambient
conditions 2
,
proceed as follows (gas equation):
007
008
009
overflow
low batt
l/min
timer
total
in
l h
WARNING
Use the medical device only within the specified
pressure range and environmental conditions.
Observe section "Technical Data". Otherwise,
the correct functioning of the medical device
may be compromised.
005
*
Standard liters per minute at 20 °C and 1013 hPa
010
l/min
NOTE
If "overflow" is shown on the display, the measure-
ment range was exceeded.
Turn thumb wheel clockwise until "overflow" disap-
pears.
When the measurement range is exceeded, the
displayed flow does not correspond to the actual
flow.
01
1
2 =
1 x T2 x p1
p2 x T1
1
= Flow displayed in L/min (20 °C/1013 hPa)
2 = Flow rate at
p2 = Ambient air pressure (hPa)
overflow
l/min
Summary of Contents for DigiFlow
Page 163: ...DigiFlow 163 164 2 low batt 174 Dr ger Dr gerService Dr gerService Dr ger Dr ger...
Page 164: ...164 DigiFlow Dr ger EMC IEC 60601 1 2 179 O2 O2 DigiFlow...
Page 176: ...176 DigiFlow CISPR 11 1 IEC 61000 3 2 IEC 61000 3 3...
Page 258: ...258 DigiFlow CISPR11 1 A IEC 61000 3 2 IEC 61000 3 3...
Page 263: ...This page intentionally left blank Instructions for Use DigiFlow 263...