84
Instructions for use Babyleo TN500 SW 1.0n
Operation
6.8.7
Setting the advanced skin temperature alarm limits
With the
Advanced skin temperature monitoring
function the alarm limits for the
difference between the central and the peripheral skin temperature can be defined
additionally.
Prerequisite:
–
The 2 skin temperature sensors are connected to the device and attached to the
patient.
Procedure:
1
Touch the
System setup...
button in the main menu bar.
2
Open the
Alarms
screen.
3
Enter the password and confirm.
4
Touch the
Skin mode
tab.
5
Select
Advanced skin temperature monitoring
>
On
and confirm.
The additional buttons for setting the skin temperature alarm limits are
displayed.
Fig. 43
Additional skin temperature alarm limits
6
Set and confirm the alarm limits for the difference between the central and the
peripheral skin temperature.
✓
The advanced skin temperature alarm limits are set.
If one of the advanced skin temperature alarm limits is switched off, the following
symbol is displayed on the main screen:
The same symbol is displayed if skin temperature monitoring is deactivated.
28483