Instructions for use Babyleo TN500 SW 1.0n
Troubleshooting
!! 212
Skin temperature low
The tolerated deviation for
central skin temperature (yel-
low) has been exceeded.
The current central skin tem-
perature is too low.
Check the settings.
Check the patient's condition.
Make sure that the sensor is
correctly attached to the
patient.
Close the hood, the side
access panels, or the hand
ports.
!! 54
Skin temperature not con-
firmed
The setting was changed,
but the change was not con-
firmed.
Adjust the setting for skin
temperature and press the
rotary knob to confirm.
!
118
Start/standby key faulty
The start/standby key has
been pressed for more than
10 seconds.
Release the start/standby
key.
Contact specialized service
personnel.
!!! 219
Temp. of convective heater
high
An internal fault occurred.
Switch off the device and let
it cool down.
Contact specialized service
personnel.
!! 17
Temp. of humidifier heater
high
An internal fault occurred.
Switch off the device and let
it cool down.
Contact specialized service
personnel.
!! 164
Temperature high in warm
air duct
The warm air duct is cov-
ered.
Make sure that the warm air
duct is not covered.
An internal fault occurred.
Contact specialized service
personnel.
!! 207
Warmer off, check patient's
condition
The radiant warmer was run-
ning at more than 30 % for
15 minutes. The radiant
warmer is off.
Check the set value.
To continue with the last
active settings for the radiant
warmer, press the "Alarm
reset" key.
!! 55
Warmer power not con-
firmed
The setting was changed,
but the change was not con-
firmed.
Adjust the setting for warmer
power and press the rotary
knob to confirm.
!
109
Warm-up completed
The set target value for
warm-up is reached.
Check the settings and adjust
them if applicable.
!
165
Water level low
The water level of the humid-
ification system is low.
Make sure that the water
supply is correctly con-
nected.
Refill the water supply.
!
110
Weaning completed
The set target value for
weaning is reached.
Check the settings and adjust
them if applicable.
Remove the patient from the
device.
Alarm pri-
ority
Alarm
Cause
Remedy
Summary of Contents for Babyleo TN500
Page 2: ...2 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Page 212: ...212 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Page 214: ...214 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Page 221: ...Instructions for use Babyleo TN500 SW 1 0n 221 This page has been left blank intentionally...