Instructions for use Babyleo TN500 SW 1.0n
Operation
6.2.1.2
Adjusting the therapy settings
Before a patient is placed in the device, the device must be warmed up.
Prerequisites:
–
The device is in operation.
–
The device has been warmed up.
–
The device is in incubator operation or radiant warmer operation.
WARNING
Risk due to incorrect settings
Devices within an environment of use can be configured differently. User-specific
alarm limits or therapy settings may be set on the devices. If the settings are
unsuitable for the patient, the patient may be put at risk.
►
Ensure that the set therapy values are suitable for the new patient.
►
Make sure that the alarm system has not been rendered ineffective by setting
the alarm limits to extreme values or deactivated by switching off the alarms.
►
Check the basic alarm settings and the patient-specific alarm settings when a
new therapy is started.
CAUTION
Risk of hypothermia
If the device has not been warmed up, the patient may become hypothermic.
►
Wait for the warm-up time of the radiant warmer.
►
In incubator operation, wait for a warm-up time of 30 minutes.
Procedure:
1
Select the desired mode in the therapy bar.
2
Set the desired temperature with the therapy control.
3
Set additional parameters (e.g., humidity).
✓
The therapy settings have been adjusted.
Summary of Contents for Babyleo TN500
Page 2: ...2 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Page 212: ...212 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Page 214: ...214 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Page 221: ...Instructions for use Babyleo TN500 SW 1 0n 221 This page has been left blank intentionally...