22
Dräger X-dock 5300/6300/6600
Basics
3.1
Special features when servicing an
X-am 8000 with PID
When using the maintenance station with an X-am 8000 with
PID, isobutylene must be set as test and calibration gas for the
PID on the X-am 8000 via the Dräger CC-Vision PC software.
Otherwise, an error message will be displayed.
Connect the test gas cylinder with isobutylene to the first
available test gas inlet to ensure that the test gas has the
highest possible quality when reaching the gas detector.
To improve the quality of the PID calibration, and to avoid that
the sensor is unnecessarily influenced by other test gases,
Dräger recommends selecting the "Flush" test option to flush
the module with fresh air after performance of a test.
Dräger recommends not storing an X-am 8000 with PID in the
module with the cover closed for a longer period of time, to
avoid drifting of the PID.
For precise measurements with the PID LC, Dräger
recommends repeating the zero adjustment on the sampling
site with the X-am 8000 using an activated carbon pre-tube.
The span calibration does not need to be repeated.
If an X-am 8000 with a PID LC has been stored for a longer
period of time in the X-dock, the calibration must be checked
prior to using the gas detector.
4
Basics
4.1
Switching the station on or off
To switch the station on:
Press and hold the key on the master for approx.
1 second.
The following information is displayed during the switch-on
process:
Software version number
To switch the station off:
Press and hold the key on the master for approx.
3 seconds.
The station switches off.
Standby mode:
Standby mode is activated after approx. 10 minutes with no
activity on the station (input via touchscreen or opening/
closing of a module cover).
When the station switches over to standby mode, any user
logged in will be logged off automatically. The user will
need to log back in again when switching back to operating
mode.
The touchscreen is switched off in standby mode.
The charging function of the X-am 125+ and X-am 8000+
modules with charging function is not affected by standby
mode. The charging process is not interrupted.
To switch to operating mode:
Tap the function key or
touch the touchscreen or
open or close a module flap.
4.2
Initial setup of station
1. Switch on the station, see chapter 4.1 on page 22.
2. Log in with the preconfigured "admin" user
(user name: admin, password: 123456), see chapter 4.5
3. Configure the test gas inlet, see chapter 4.6 on page 24.
4. Change language if necessary:
a. Select
> System configuration > Language
.
b. Select desired language.
c. Confirm selection with
OK
.
5. Set date and time where necessary:
a. Select
> System configuration > Date & time
.
b. Make the desired settings.
c. Confirm settings with
OK
.
NOTICE
If no action has occurred for 10 minutes, any user
logged in will be logged off automatically. After
45 minutes, the screensaver will be activated.