42
Fault messages
Special symbol » « and
displayed numerical code:
Cause
Remedy
01
Dräger X-am 5000 with alkali supply unit. Insert Dräger X-am 5000 with battery
power pack.
02
Communication with Dräger X-zone 5000
interrupted.
Check IR interface on Dräger X-zone
5000 and on Dräger X-am 5000.
03
Communication error with battery
controller Dräger X-zone 5000.
Contact Dräger.
04
Main battery alarm Dräger X-am 5000.
Check charging contacts on
Dräger X-zone 5000 and on
Dräger X-am 5000.
05
Battery pre-alarm Dräger X-am 5000.
Check charging contacts on
Dräger X-zone 5000 and on
Dräger X-am 5000.
06
Dräger X-am 5000 charging current too
low.
Check charging contacts on
Dräger X-zone 5000 and on
Dräger X-am 5000.
07
Holder (pump) detected, but no pump
fitted.
Use holder for diffusion mode.
08
Flow fault
Check intake hose.
09
Holder status change (pump) during
operation.
Check that the holder (pump) is securely
positioned.
10
Check sum error program code
Contact Dräger.
11
Check sum error operating parameter
Contact Dräger.
12
Check sum error operating parameter
Contact Dräger.
13
Check sum error operating parameter
Contact Dräger.
14
Working memory test error
Contact Dräger.
15
Faulty ADC conversion.
Contact Dräger.
16
No contact with switch box in grouping
mode.
Check wireless connection to switch box.
17
Charging electronics faulty.
Contact Dräger.
18
Battery completely discharged.
Charge Dräger X-zone 5000.
19
Battery main alarm of the
Dräger X-zone 5000.
Charge Dräger X-zone 5000.
20
Battery pre-alarm of the
Dräger X-zone 5000.
Charge Dräger X-zone 5000.
21
Dräger X-am 5000 device error.
Check Dräger X-am 5000.
22
Dräger X-am 5000 alarm pattern faulty.
Contact Dräger.
23 - 28
-
-
29
Battery main alarm
Charge Dräger X-zone 5000.
30
Battery completely discharged.
Charge Dräger X-zone 5000.
31
Charging electronics faulty.
Contact Dräger.
32
Communication error with battery
controller Dräger X-zone 5000.
Contact Dräger.