30
Info window for Dräger X-am 5000:
1
Station number in grouping mode:
The station number is not currently supported by the
Dräger X-zone 5000.
2
Wireless status: ON/OFF:
Displays whether the wireless function of the Dräger
X-zone 5000 is switched on or off.
The wireless function can be switched on and off
using the Dräger CC-Vision PC software (refer to
“Configuring the device” on page 39).
3
Network number:
If independent wireless networks are required,
different network numbers must be assigned (refer to
“Establishing a wireless connection” on page 32).
The network number can be set using the Dräger
CC-Vision PC software (refer to “Configuring the
device” on page 39).
4
Grouping ID:
The grouping ID is not currently supported by the
Dräger X-zone 5000.
5
X-zone mode symbol
Displays whether or not there is a connection
between the Dräger X-zone 5000 and the Dräger X-
am 5000.
– If no key is pressed for 10 seconds, the Dräger X-am
5000 returns automatically to detection mode.
Switching off the device
STANDBY mode
z
Switch off the Dräger X-am 5000 in the Dräger X-
zone 5000 in accordance with the instructions for
use.
– The visual and the audible alarm will be activated for
a short time before the Dräger X-zone 5000 switches
off.
– The Dräger X-zone 5000 switches to the STANDBY
mode.
OFF mode
z
Release the lock at the holder.
z
Remove the holder.
z
If necessary, remove the Dräger X-am 5000 from the
device receptacle.
z
Press the key and the key on the
Dräger X-zone 5000 and hold down for approx.
3 seconds.
– The battery LED goes off.
– The Dräger X-zone 5000 is switched off
(OFF mode).
Power supply for the Dräger X-am 5000
with NiMH battery in the Dräger X-zone
5000:
Carrying out a bump test
The bump test can be carried out in two ways.
Bump test for the Dräger X-am 5000:
z
Perform a bump test in accordance with the
Instructions for Use for the gas detection instrument
before using with the Dräger X-zone 5000.
Bump test on the Dräger X-am 5000 in combination
with the Dräger X-zone 5000:
In diffusion mode
z
Switch the Dräger X-zone 5000 on (refer to
“Switching on the instrument” on page 29).
CAUTION
When switching off the Dräger X-zone 5000, Dräger
recommends wearing hearing protection or using the
alarm-attenuation ring (order no. 83 20 110) as the
acoustic alarm is activated for a brief period.
025331
12.eps
RF:
NET:
P\N:
ON
--
005
FFFF
1
2
3
4
5
Device mode
Power supply for
the Dräger X-am 5000
Dräger X-zone 5000: ON mode
Dräger X-am 5000:
Switched on
Dräger X-am 5000 is
permanently supplied
with power.
Dräger X-zone 5000: STANDBY mode
Dräger X-am 5000:
Switched off
Dräger X-am 5000 is
supplied with power
via trickle charging.
Dräger X-zone 5000: OFF mode
Dräger X-am 5000:
Switched off
Dräger X-am 5000 is
not supplied with power.
CAUTION
A bump test must be carried out on every device
before use.
OK
027331
12.eps
1
2