
16
Instructions for use Humidifier Nebulizer
English
English
Carry out machine cleaning and disinfection
1
Strictly observe Instructions for Use of washer-
disinfector.
2
Position items so that all interior spaces are
completely flushed and water can drain off freely.
3
Use suitable cleaning agent.
4
Select suitable program (preferably anesthesia
program).
– Cleaning must be carried out at 40 °C to
60 °C (104 °F to 140 °F) for at least 5
minutes.
– Thermal disinfection must be carried out at
80 °C to 95°C (176 °F to 203°F) and with
corresponding contact time.
5
Carry out final rinsing with deionized water.
6
Immediately remove items from the washer-
disinfector.
7
Check parts for visible dirt and damage. If
necessary, repeat program or carry out manual
cleaning and disinfection.
8
Allow items to dry thoroughly.
Sterilization
During sterilization, living microorganisms are
removed from semi-critical medical devices.
Residual water inside the components is also dried
out.
Only sterilize cleaned and disinfected items.
Use a vacuum steam sterilizer (in accordance with
DIN EN 285), preferably with fractional vacuum.
Visual inspection
Inspect all items for damage and wear, e.g.
cracking, embrittlement or pronounced
hardening, and residual soiling.
Disposal of the Medical Device
When disposing of the medical device:
Consult the relevant waste disposal company for
appropriate disposal.
Observe the applicable laws and regulations.
Service
Safety information
CAUTION
Even accessories designed to be reused have a
limited service life. Handling and reprocessing can
increase wear and markedly shorten service life
(e.g., disinfectant residues can attack the material
more intensely during autoclaving). If signs of wear
become visible, such as cracks, deformation,
discoloration, peeling, etc., affected accessories
must be replaced.
WARNING
Risk due to inappropriately reprocessed
products
The product may be contaminated with
infectious agents.
Before service is performed and before the
product is sent back for repair, reprocess the
product in accordance with the chapter
"Reprocessing".
WARNING
Risk if service is not performed regularly
Wear and material fatigue of the components
may lead to device failure and malfunctions.
Perform service at the specified intervals.
WARNING
Risk if service is not performed properly
Personal injury and property damage may
occur if service is not performed properly.
Service must be performed by those target
groups that are assigned to the particular
measure.
Summary of Contents for 2M85835
Page 63: ...63 Dr ger Dr gerService...
Page 64: ...64 Dr ger O2 O2 DigiFlow O2 O2 O2...
Page 68: ...68 EN ISO 15883 1 2 3 4 40 C 60 C 104 F 140 F 5 80 C 95 C 176 F 203 F 5 6 7 8 DIN EN 285...
Page 69: ...69 Dr ger Dr gerService Dr ger O2 O2 DigiFlow O2...
Page 132: ...132 Dr ger Dr gerService...
Page 133: ...133 Dr ger O2 O2 DigiFlow O2 O2 A B C D O E F G H O2 004 A B C D E F G H...
Page 135: ...135 A O2 O2 DigiFlow B O2 O2 O2 DigiFlow A O2 O2 DigiFlow B C D O2 O2 DigiFlow 003 A B D C...
Page 137: ...137 Dr ger Dr gerService Dr ger Me O2 O2 DigiFlow...