1-4
1-4-3 How to GROUND
・
For safe operation, be sure to ground the ground terminal of power supply at one point
on the ground.
・
If the Load frame is not grounded
・
If the Load frame is grounded
・
Make sure to connect the GROUND terminal correctly as shown in Fig.1.2 and Fig.1.3. Poor grounding
is dangerous and may cause an electric shock or damage the power supply.
・
When load short-circuit or discharge is forecasted, make the GROUND cable thicker and shorter.
Ground at one point
Fig.1.3
Fig.1.2
output cable
load frame
ground
ground terminal
rack
–
mount case
Wires should be :
・more than AWG10
・less than about 2m
・not bared
Ground at one point
load
load
output cable
ground terminal
load frame
rack
–
mount case
Wires should be :
・more than AWG10
・less than about 2m
・not bared
!
MUST
!
MUST
!
MUST
!
MUST
!
MUST
CAUTION
1 I n t r o d u c t i o n