4 Advanced Operation
4-8 Output current monitor, Output voltage monitor
4-8-1 Output current monitor (L1)
10V voltage is output at the maximum output current. The output impedance is 1kΩ.
The monitor output polarity is positive regardless of the high voltage output polarity.
4-8-2 Output voltage monitor (L2)
10V voltage is output at the maximum output voltage. The output impedance is 1kΩ.
The monitor output polarity is positive regardless of the high voltage output polarity.
4-9 HV status signal
When the OUTPUT ON/OFF switch is ON, the internally built-in relay contact is ON.
The voltage is 30Vdc MAX and allowable current is 100mA MAX
between pin 17 and pin 19 when contact is open
TB1