Mach M5
Dr. Mach
Lamps and Engineering
1. Safety instructions
Pay attention to the instructions for use when handling the lamp.
WARNING:
This device has not been designed for use in potentially explosive areas.
According to the Medical Device Regulation the lamp is classified under class I.
Store the OT-lamp in its package for at least 24 hours in the respective room before
mounting, in order to equal temperature differences.
Please read the instructions for use carefully to make the most of your lighting
system and to avoid any damages to the device.
The lamps may only be repaired and special assembly work may only be carried
out on the reflector or sockets by ourselves or a company that has been expressly
authorised by us.
The manufacturer can only be made responsible for the safety of the lamp if re-
pairs and alterations are carried out by the manufacturer himself or a company that
guarantees to observe the safety regulations.
The manufacturer cannot be made liable for personal or material damages if the
lamp is operated inexpediently or incorrectly or used for purposes other than
those for which it is intended.
The lamp is to be dismantled from the spring arm in reverse order to its assembly.
This may only be carried out after the spring arm has been adjusted in height at
horizontal position since the arm is under spring tension and can bounce up.
Make sure that the lamp is in perfect working order before every use.
Attention, external transformer!
The lamp works only with an external transformer 300VA.
The external transformer (Order No. 6701 0208) must be tested and validated ac-
cording to IEC 60601-1.
If the external transformer is not tested and validated according to IEC 60601-1, it
is not allowed to use it with Dr. Mach OT-lamps.
59270001
Edition 06
28.09.2004 / Bak page 4/26