Mach M5

 

Dr. Mach

Lamps and Engineering

 

 

Instructions for use 

 

MACH M5 

 

 
 
 

 

 
 
 
 
 
 
 
 
 
 

Mach M5 ceiling lamp for room heights up to 2,8m, F- / F/H / DF- / DF/H-model 
Mach M5 ceiling lamp for room heights between 2,8m – 3m,

 

F- / F/H / DF- / DF/H-model 

 
 
 

 

Dr. Mach

 

GmbH u. Co., Floßmannstrasse 28, D-85560 Ebersberg 

Tel.: +49 (0)8092 2093 0, Fax +49 (0)8092 2093 50

 

Internet: 

www.dr-mach.com

, E-Mail: 

[email protected]

 

59270001 

Edition 06 

28.09.2004 / Bak  page 1/26

 

Summary of Contents for Mach M5

Page 1: ...ights up to 2 8m F F H DF DF H model Mach M5 ceiling lamp for room heights between 2 8m 3m F F H DF DF H model Dr Mach GmbH u Co Floßmannstrasse 28 D 85560 Ebersberg Tel 49 0 8092 2093 0 Fax 49 0 8092 2093 50 Internet www dr mach com E Mail info dr mach de 59270001 Edition 06 28 09 2004 Bak page 1 26 ...

Page 2: ...page 7 2 6 Adjusting the light intensity page 8 2 7 Positioning page 9 3 Cleaning page 10 3 1 Sterilisable handle page 10 3 2 Lamp housing dispersing lens and support system page 12 4 Maintenance page 12 4 1 Adjustments at the lamp head page 12 4 2 Changing of spare parts page 13 4 2 1 Changing the halogen bulbs page 13 4 2 2 Changing the filter disk page 15 4 2 3 Changing the dispersing lens page...

Page 3: ...lue of Ra 96 In other words the colours are reproduced naturally and in high con trast The wound area is shown in a comfortable light The different reds of a wound area can now be recognised very precisely For the doctor or surgeon this means a considerable improvement in the recognition of details in the wound area The R96 lighting system uses computer optimised cold light filters that reduce bot...

Page 4: ...amp if re pairs and alterations are carried out by the manufacturer himself or a company that guarantees to observe the safety regulations The manufacturer cannot be made liable for personal or material damages if the lamp is operated inexpediently or incorrectly or used for purposes other than those for which it is intended The lamp is to be dismantled from the spring arm in reverse order to its ...

Page 5: ...y swiveling them NOTE This feature is activated by turning the sterilisable handle 2 1 2 Focusing DF model The DF model offers a second additional fa cility to further adjust the illuminated field to an extra small or extra large size depending on the specific requirements by focusing through moving the bulbs up or down NOTE This feature is activated by turning the ring at the top of the sterilisa...

Page 6: ...erging the light fields and focusing Duo Focus system for DF and DF H models DF and DF H models also have a focusing func tion Duo Focus The combination of merging the light fields and focus ing Duo Focus system enables the ideal adjust ment of the light field to the wound area in a range from 8cm to 35cm diameter Such a large range of adjustment is not possible with conventional tech nique To act...

Page 7: ... done by moving the bulbs inside the re flectors up or down This way you can further adjust the illuminated field to an extra small or extra large size depending on the specific requirements This feature is activated by turning the ring 2 at the top of the sterilisable handle FOCUSING MERGING OF LIGHT FIELDS Ring 2 Handle 1 Always activate merging of light fields first by turning the handle 1 then...

Page 8: ...adjustment is done by the key pad at the cardanic bow The intensity can be ad justed continuosly between 50 and 100 Press push button 2 to increase the light intensity and press push button 3 to reduce the light inten sity The set light intensity is shown by LED display 1 For other light models we recommend a dimmer in the supply line 1 2 3 59270001 Edition 06 28 09 2004 Bak page 8 26 ...

Page 9: ...he working distance for the F lamps is 0 7m to 1 4m The working distance for the DF lamps is 0 6m to 1 5m The handle is supplied with a detectable mechanical catching When the handle catches in this position the lamp is optimally set for a working distance of 1 meter 1 2 1 4m 1 5m 1m 0 7m 0 6m 0 Two or more lamps can be used for an intensive illumination of large operation fields It is also possib...

Page 10: ...cific processes are used for cleaning disinfection and that the vali dated parameters are complied with in every cycle In addition the hospital clinic hygiene regulations must be observed Cleaning disinfection Cleaning and disinfection must be carried out immediately after use A mechanised process disinfector should be used for cleaning disinfection The efficiency of the proc ess used must be reco...

Page 11: ...table sterilisation pack one way sterilisation pack e g foil paper sterili sation bags single or double pack in accordance with DIN EN 868 ISO 11607 for steam sterilisation and then sterilised Use only the sterilisation procedure listed below for sterilisation Other sterilisation procedures e g ethyl ene oxide formaldehyde and low temperature plasma sterilisation are not permissible Steam sterilis...

Page 12: ... adjusted The lamps and suspension fixtures have been designed and built so that regular maintenance intervals are not necessary In order to keep the system easy running throughout its life span we recommend that the hinges should be greased once a year with acid free grease Attention Before dismounting the lamp set the height adjustment of the spring arm to horizontal position see mounting instru...

Page 13: ...n bulbs in a lamp at the same time to keep mainte nance work to a minimum To change the halogen bulbs proceed as follows Turn off the lamp ATTENTION Parts of the housing and the halogen bulb may be very hot immediately after use Turn the lamp so the light outlets point to the ceiling Turn screw 1 anticlockwise with an appropriate screw driver Remark In case there is no screw driver available you c...

Page 14: ...with your fingers Always use the original packaging or a clean cloth Carefully insert the new halogen bulb 22 8V 77W in the socket Tilt the shadower 4 in its original position Mount the optic ring 2 in reverse order anti clockwise rotation For mounting position the optic ring 2 so the toothed wheel segment 6 of the ring and the toothed wheel 7 in the lower housing part work into each other Fix the...

Page 15: ...ge the filter disk proceed as follows Turn off the lamp ATTENTION Parts of the housing and the halogen bulb may be very hot immediately af ter use Turn the lamp so the light outlets point to the ceiling Turn screw 1 anticlockwise with an appropriate screw driver Remark In case there is no screw driver available you can use a coin By turning the screw 1 anticlockwise the optic ring 2 turns clockwis...

Page 16: ...ull it off the other three retaining angles Mount the new filter disk in reverse order ATTENTION Only use filter disks that have been cleaned properly see chapter 3 2 Mount the optic ring 2 in reverse order anti clockwise rotation For mounting position the optic ring 2 so the toothed wheel segment 5 of the ring and the toothed wheel 6 in the lower housing part work into each other Fix the optic ri...

Page 17: ...hing base to preserve its surface To change the dispersing lens proceed as follows Turn off the lamp ATTENTION Parts of the housing and the halogen bulb may be very hot immediately after use Turn the lamp so the light outlets point to the ceiling Turn screw 1 anticlockwise with an appropriate screw driver Remark In case there is no screw driver available you can use a coin By turning the screw 1 a...

Page 18: ...ly use dispersion lenses that have been cleaned properly see chap ter 3 2 Mount the retaining ring 6 the retaining an gles 5 the filter disk 4 in reverse order and fix the retaining angles with the six screws 3 Mount the optic ring 2 in reverse order anti clockwise rotation For mounting position the optic ring 2 so the toothed wheel segment 8 of the ring and the toothed wheel 9 in the lower housin...

Page 19: ...n the ground of a normed tube with one shadower 125 000 Lux Light intensity on the ground of a normed tube with two shadowers 84 600 Lux Illumination depth 715 mm Colour rendering index Ra at 4300 Kelvin 96 Colour rendering index R9 at 4300 Kelvin 90 Focusable light field size 17 35 cm Colour temperature Kelvin 4300 K Radiation intensity in field at 100 000 Lux 380 W m 2 Temperature increase in he...

Page 20: ...sure 700 hPa 1060 hPa Transport storage Min Max Temperature 10 C 50 C Relative athmospheric humidity 20 90 Air pressure 700 hPa 1060 hPa 5 4 General remarks When using more than one OT lamp at the same time OT lamp combinations please make sure that the total radiation intensity in field is less than 1000 W m2 to avoid further temperature increase in the wound field When installing an OT lamp its ...

Page 21: ...001 1996 and DIN EN ISO 13485 2001 7 Disposal The OT lamp doesn t contain any dangerous goods The components of the OT lamp should be properly disposed at the end of its shelf life Make sure that the materials are carefully separated The electrical conducting boards should be submitted to an appropriate recycling proceeding The rest of the components should be disposed according to the contained m...

Page 22: ...Mach M5 Dr Mach Lamps and Engineering 8 Spare parts 4 5 3 15 16 25 6 7 23 24 1 2 8 20 21 22 19 18 17 9 10 11 12 13 14 40 59270001 Edition 06 28 09 2004 Bak page 22 26 ...

Page 23: ...Mach M5 Dr Mach Lamps and Engineering 27 26 28 30 31 29 32 35 36 34 33 59270001 Edition 06 28 09 2004 Bak page 23 26 ...

Page 24: ...d OPLSD II 67250110 F DF H lamps with Bowden control 22 1 Board OPLSD IV 67250112 F DF lamps without light intensity control 23 1 Lamp bow 27101001 24 1 Lamp bow for central spring arm 27100001 For low room height 25 1 Cardan bow 27111001 26 3 Halogen bulb 22 8V 75W 67100207 27 3 Ceramic lamp socket with cable 67320007 28 3 Shadower 27061001 29 3 Screw 27089203 30 1 Keypad holder incl keypad F H l...

Page 25: ...os 3 secondary side 3 1 2 Connect the transformer to earth Use the provided clamps pos 1 at the transformer respectively at the transformer plate See also the explanations on page 19 46 of the mounting instructions Ceiling and wall at tachment no 59500001 edition 05 Overview of the possible settings of the primary voltage Setting for115V Setting for 132V 59270001 Edition 06 28 09 2004 Bak page 25 ...

Page 26: ...Mach M5 Dr Mach Lamps and Engineering 59270001 Edition 06 28 09 2004 Bak page 26 26 Setting for 230V Setting for 247V ...