6
and dry cloth and leave to dry naturally in a clean place. To sterilize, cover the medication container with a cold sterilizing
solution, following the instructions on the sterilizer pack. It is recommended to use one personal medicine cup for each
patient, to be used for 6 months or 120 treatments. Do not use boiling water to clean it. Carefully wash masks and hoses
with disinfectant solution only.
CONTAMINATION BY MICROBES
In case of pathologies of microbial or viral origin, where infection is possible, the device and its accessories should not
be shared with other patients.
Troubleshooting
PROBLEM FACTOR
SOLUTION
Too much noise No filter in the unit.
Put a filter in the unit.
The nebulizer
does not run
Compressor switch is off.
Turn the unit on.
Nebulizer was not cleaned after
its previous use.
Clean the unit.
Air hose is bent.
Remove any bends or kinks in the air hose.
The filter is clogged.
Replace the filter.
No medication is left.
Add the appropriate amount of medication prescribed
by your physician to the cup.
The cap is not attached properly. Reattach the cap.
Water drops
have formed on
the air hose.
The medication was overfilled.
Add appropriate amount of medication to cup,
connect the air hose to the air compressor, and turn
the unit on.
The nebulizer was washed but
not dried.
Open and close the nebulizer repeatedly with your
finger until the drops disappear.
TECHNICAL SPECIFICATIONS
Compressor nebulizer machine. Complies with the requirements of European Standard EN 13544-1. Rotating pressure
piston, without lubrication. Medical device of class IIA according to Directive 93/42/EEC. Linear medicine cup.
Model
Turbo Lex
Adapter input
100 - 240 V ~ , 50 / 60 Hz, max 0.5 A
Main unit Input
12V
Noise Level
≤ 60 dB
Max capacity of medication cup
8 ml
Particle Size
0.5 to 6 um
MMAD
4.0 ± 20% μm
Nebulizing rate
≥ 0.24 ml/min
Operation mode
30 minutes on, 30 minutes off
Operation temperature, humidity and atmospheric
pressure
10°C ~ 40°C, ≤ 85 % relative humidity
860 ~ 1060 hPa
Transport and storage temperature
humidity and atmospheric pressure
-10°C ~ 40°C, ≤ 85 % relative humidity
860 ~ 1060 hPa
Pollution degrees
Degrees 2
Overvoltage category
Category II
Maximum operational altitude, m
2000
NOTE:
If the device is not functioning correctly in spite of inspections effected, contact the ТМ Dr. Frei
®
service
center in your country.
Summary of Contents for TURBO LEX
Page 1: ...MODEL TURBO LEX...
Page 9: ...RU 9 36...
Page 10: ...10 Wi Fi 1 2 3 4 5 6 5 7 IIa 1 2 3 2 2 3 3 9 6 8 5 4 1 9 7...
Page 11: ...RU 11 4 5 B 6 C 7 9 D 8 E 9 ON 10 45 100 I 9 6 120 A B C D E H I...
Page 13: ...RU 13 0413 BF I ON O OFF IP 21 12 Serial number 2 Dr Frei Dr Frei IEC60601 1 2...
Page 15: ...UA 15 i 36...
Page 17: ...UA 17 45 30 I 9 C 5 10 6 120 ON 7 8 9 Dr Frei H I...
Page 26: ...26 26 26 26 26 26 26 27 28 28 28 28 28 29 29 30 31 31...
Page 27: ...BG 27 36 c Wi Fi 1 2 3 4 5 6 5 7 A 2 2 3 3 9 6 8 5 4 1 9 7...
Page 28: ...28 IIa 1 2 3 4 5 B 6 C 7 8 8 D 8 E 9 ON 10 45 100 H I 9 H I A B C D E A B C D E...
Page 30: ...30 10 C 40 C 85 860 1060 2 II 2000 5 160 x 110 x 62 0 275 2000 8 0413 93 42 BF ON OFF IP 21 12...