58
Confused
result of mea-
surement
Measurement not
continuous
Measure temperature
after 3 minutes
Sensing head
is dirty
Use a cotton swab or
soft cloth and wipe it
gently with alcohol
Temperature
sensing head
is damaged or
scratched
Please contact agent
to repair. (Must be
under warranty)
The ear has
foreign matter
within it.
Consult a doctor
Measurement
done incorrectly
Please read the
manual carefully
Display shows
“HH“
Indicates the
system detects
the ambient
temperature ex-
ceeds the normal
operational range
of the device
The measurement
should be done in an
environment between
15°C to 20°C and
the device should
be stabilized for 30
minutes in such an
environment
Display shows
“LI“
Indicates the
system detects
the ambient
temperature ex-
ceeds the normal
operational range
of the device
The measurement
should be done in an
environment between
15°C to 20°C and
the device should
be stabilized for 30
minutes in such an
environment
Display shows
“Hi“
Indicates that the target temperature is
above the measurement limit of the device
Display shows
“Lo“
Indicates that the target temperature is
above the measurment limit of the device
MAINTENANCE
Cleaning and Care
• The detecting sensor is not waterproof. Please wipe with a