10. Prior to transistor test you should make sure that gauging probes have been disconnected from
another gauged circuit.
11. Prior to measurement of resistance, circuit continuity and capacitor’s capacity the capacitances
should be discharged and all power supply sources disconnected.
12. Particular caution should be taken by measurements, which exceed DC 60 V or AC 30 V rms.
13. Prior to commencement of measurement the function range switch should be set in an adequate
position. Shifting switch during the measurement may cause damage to the device.
14. Gauging probes should be removed from gauging slots prior to every change of measured parame
-
ters.
15. The device should not be used, or kept, in high air humidity or temperature conditions, in strong
electromagnetic field or in an explosive or flammable environment. Such conditions may influence
gauging results and create an electric shock hazard.
16. The device should not be used when the display indicates exhausted battery. Low battery level may
lead to faulty gauging indications.
17. Make sure that the device is switched off before batteries are replaced.
18. In case device is not used for a longer period of time the battery should be removed in order to avoid
spilling of electrolyte.
19. Device should be used and stored in a place inaccessible for children.
20. Device is designed to be used indoors, in room temperature.
21. Device should not be used directly after it is relocated from a room with high air humidity, excessive
-
ly low or high temperature conditions.
Technical specification
Universal clamp multimeter
measurements: AC amperage, AC/DC voltage, resistance, diode test, cir
-
cuit continuity test 3.5 digit LCD display
auto ranging
indicators: polarity, over range, exhausted
battery
buzzer
HOLD – function remembering the last measurement auto power off (after 15 min
standby)
overload protection
power supply: 2 x AAA (included) number of readouts: 2 readouts
per second
operating conditions: from 0 ºC to 40 ºC, air humidity: < 80% storage conditions:
from – 10 ºC to +50 ºC, air humidity < 75%
Measurement accuracy for particular gauging values is given for 1 year from calibration date and for
an operating temperature of 23ºC +/– 5ºC and air humidity of 75%. Accuracy: +/–% of indicated
value +/– number of least significant digits.
Safety category
CAt I
– gauging category CAT I defines safety requirements for measurements in devices which are not
connected directly to a low voltage network, such as batteries, accumulators, flashlights.
CAt II
– gauging category CAT II defines safety requirements for measurements carried out in devices
directly connected to a low voltage network, such as home appliances, office equipment or workshop
equipment.
Multimeter should not be used for measurements of devices defined by CAT III and CAT IV.