Ultra Evac Technical Manual
18
2/8/2013
GD-99013 Rev. D
De-install/Disconnect Instructions
Please contact DMS at 1-888-466-6633 prior to de-installation or re-installation of an evac.
If a Customer de-installs or re-installs an evac, the guidelines listed below must be followed.
Always wear personal protective equipment (PPE), including eye protection and latex gloves, to
prevent contact with potentially infectious material.
Always unplug the unit from the wall electrical receptacle before moving equipment.
Wipe down the unit with an approved contact disinfectant after each use.
Equipment Storage
Temperature range for proper storage is -20
o
C to 40
o
C. If unit is to be taken out of service and stored for a
prolonged period of time, evacuate water from the hot and cold water manifolds, as well as removing the
bleach and enzyme solution from their chemical tanks regardless of temperature within storage facility.
Operation
Storage and Transportation
Temperature
Relative
Humidity
Atmospheric
Pressure
Figure 10
– Environmental Conditions