Ultra Evac Technical Manual
20
2/8/2013
GD-99013 Rev. D
Routine Maintenance
Front Panel Removal
Call DMS at 1-888-466-6633 prior to removing or servicing any components that require the front panel to
be removed. Removal of the front panel will allow access to the electrical components underneath. Once
removed the possibility of electrical shock is present. Use caution when performing any service to these
components. To take of the front panel, remove the two screws on either side of the green top and raise
the top. Remove the two screws on either side of the front panel near the bottom of the unit at the back
edges of the panel. Lift the panel up and out to remove. Note: This panel has a ground wire attached to it.
Inspections
Regularly visually inspect the unit for any signs of damaged or leaking components. If any problems are
found, contact Customer Service at 1-888-466-6633 to report findings.
Replacement Items
Power Supply
There is not a substitute power supply replacement. The power supply must be replaced with the following:
Manufacturer: Condor
Model #: GLM75 Medical
Contact DMS for replacement parts at 1-888-466-6633.
Replacement Fuses
Two sets of replacement fuses are located on the main circuit board. Fuse specifications are as follows:
Fuse replacement for the main ICB power –
250V 5A time delayed fuse.
Fuse replacement for the drain pump –
250V 8A time delayed fuse.
Power Cord Replacement
Contact DMS for replacement only (see components part list), Procedure for replacement (correct
connection & anchoring) are provided with the replacement power cord.