Error Code List
After an error occurred, clear the error of the PC and the data first. Shut down the AS-410 power source and the PC; then restart. If the
error repeats after restarting the AS-410 initialization and the LK-400 application, the sample file data may have broken down or have a
failure. In such case, operate the slide preparation using a sample which has a good result in the past, and check the data contents.
Error
code
On-screen message
Possible cause
Solution
9000
<Error Message>
Unit: Remote PC
Symptom:Communication
error
Retry:
Communication failure
with the remote PC.
・
Check if the specified sample block is set
correctly; set the specified sample block
correctly.
・
Restart the LK-400 application of the PC,
and log in again.
・
Turn OFF the System power first; then start
the initialization.
・
Take out the LAN cable in case if the cable is
loosely inserted; then plug it in tightly.
9001
<Error Message>
Unit: Remote PC
Symptom:Communication
error
The specified sample
information failure from
LK-400.
9002
<Error Message>
Unit: Remote PC
Symptom:
Communication error
%PST receiving failure.
Response failure of the
start-up verification.
・
Response failure of the start-up verification.
・
Check the LK-400 screen error.
・
Restart the LK-400 application of the PC;
and log in again.
・
Shut down the Main Unit power and the PC;
restart the initialization of the Main Unit and
the
LK-400 Software. Then retry section
preparation
with the Auto Sectioning Mode or the
Manual Input Mode.
9003
<Error Message>
Unit: Remote PC
Symptom:
Communication error
%PBR receiving error.
Error of the requested
numbers of slide
preparation condition.
Ensure the numbers of the sample block and
the slide preparation condition is matched;
ensure the setting is within the tolerance
value.
1 to 96 (24):
①
Sample block numbers (*1).
0 to 400 (100):
②
Numbers of the sectioning
condition (*2).
9004
<Error Message>
Unit: Remote PC
Symptom:
Communication error
%PBI receiving error.
Errors of the slide
preparation condition
and the slide print data
(during Auto Sectioning
Mode)
・
Ensure the slide preparation condition data
of the specified sample block is normal; also,
ensure the setting is within the tolerance
value.
1 to 96 (24):
①
Sample block numbers (*1)
0 to 30:
②
Slide preparation condition
numbers.
1 to 400 (100):
③
Sectioning numbers (*2)
3 to 8:
④
Sectioning thickness
5 characters:
⑤
Slide type
9006
<Error Message>
Unit: Remote PC
Symptom:
Communication error
%PBL receiving error.
Reference error of the
sample block number.
Ensure the number of the specified sample
block and the reference numbers are correct.
①
Number of the specified sample block.
②
Sample block reference number for each.
9007
<Error Message>
Unit: Remote PC
Symptom:
Communication error
%PSB receiving error.
Errors of the slide
preparation condition
and the slide print data
(during Manual Input
Mode)
Ensure the slide preparation condition data of
the specified sample block is normal; also,
ensure the setting is within the tolerance
value.
1 to 96 (24):
①
Sample block number (*1).
0 to 400 (100):
②
Number of the sectioning
condition (*2).
9008
<Error Message>
Unit: Remote PC
Symptom:
Communication error
%PSI receiving error.
Errors of the requested
slide preparation
condition and the slide
print data (with a
Ensure the slide preparation condition data of
the specified sample block is normal; also,
ensure the setting is within the tolerance
value.
1 to 96 (24):
①
Sample block number (*1)
0 to 30:
②
Reference number of the
Summary of Contents for LK-400