The 820MO system has a build-in pH meter and a pH-meter
electrode plug-in port marked PH on the backside of the 820M
interface.
The pH electrode can be ordered at DMT by contacting your sales
representative or emailing sales@dmt.dk.
The pH calibration procedure is listed in 4 individual steps and needs
to be performed one at a time.
Step 1 - Step 1 involves cleaning the pH electrode and the temperature
probe with double distilled water. When ready, Press NEXT STEP.
Step 2 - Place the pH electrode and temperature probe in the high
buffer solution (here pH 7) and turn on stirring of the high buffer
solution. When the relative pH output in the blue line is stable, go to
NEXT STEP.
Step 3 - Place the pH electrode and temperature probe in the low
buffer solution (here pH 4) and turn on stirring of the low buffer
solution. When the relative pH output in the blue line is stable, go
to NEXT STEP.
19
NOTE: BEFORE THE PH CALIBRATION IS PERFORMED
BE SURE TO SELECT THE WAY THE PH ELECTRODE IS
TO BE USED. SEE THE SUB- MENU PH SET-UP UNDER
INTERFACE SETTINGS.
LOGIN CODE TO pH
Call DMT A/S to get access code.
System ID number. 1056
CLR
ENTER
0
1
2
3
4
5
6
7
8
9
Step no.:
Follow the pH calibration
procedure in the User Manual.
NEXT STEP
1
2
3
4
Clean the pH electrode and the
pH CALIBRATION
temperature probe.
When ready go to next step.