SETTINGS MENU
The “Settings Menu” contains several sub-menus that can be
accessed to change functional aspects of the Interface. These sub-
menus include:
I.
Force Calibration
II.
pH Calibration
III.
Force REC. OUTPUT
IV.
Measurement Range
V.
Interface Settings
1.
Force Calibration
Entering the FORCE CALIBRATION sub-menu begins the
transducer calibration procedure. Begin the calibration procedure
by pressing “FORCE CALIBRATION” to enter the sub-menu. The
sub-menu will list all 4 chambers for calibration.
(For more information and tips, please refer to Section 4.1-
Calibration of the Force Transducer)
To begin the calibration, press
“SELECT” for the chamber which
calibration will be performed on. The text for the chamber to be
calibrated will turn blue. Pressing “ENTER” will enter the 6-step
procedure for calibrating the force transducer on the desired
chamber.
The calibration procedure is listed in 6 individual steps and needs to
be performed for each channel or transducer when calibrating the
system.