Chapter 3 36
•
:
touch the
to add another calibration point
•
:
touch the
to finalize setup of the new user
material with just one calibration point
When a new material is created using one point calibration only
displacement for the selected table (base material) is generated. This is
known as a grade 0 (zero) ecuation and could be insufficient to obtain
reliable readings. In this case a second test block of the same material
but a diferent hardness value should be used to generate another
calibration point.
Select "
Other
" by touching the
to create another calibration
point.
Figure 3.27: Second point for the user material
To enter a second calibration point follow the same steps used to
enter the first calibration point.
A new user material can be calibrated with up to 4 different
calibration test blocks or samples.
Once the material has been created it will appear on the user
materials menu as selected (figure 3.22) meaning that the unit has
been calibrated for the new material. The new material will also be
displayed in the materials menu.