55 Appendix
Choose the UCI probe that is best suited for your application and take
all of the following into consideration.
Unit
Material
Steel
Aluminum
Copper
Brass
HV
100~940
44~189
40~130
45~196
HB
95~500
40~160
75~120
42~169
HRB
55~100
28~91
2~67
10~93
HRC
20~68
—
—
—
HS
32~97
—
—
—
MPa
320~2190
—
—
—
The following are minimum test piece requirements that need to be
met in order to obtain precise and repetitive measurements.
QHUS10
(10N / 1kgf)
QHUS50
(50N / 5kgf)
Roughness (Ra)
1,5 µm
2,5 µm
Curvatura Radio
>5 mm
Weight
>0,1 Kg
Thickness
<1 mm
Don’t measure of a previous indentation or in highly corroded parts.
The test piece surface must be clean, dry and free of grease.