BA FlexxPump 125 - 24 V_20181121_EN de 201811 31
External Control
7.2 Output Signals (PIN 4)
This chapter describes the signals which the FlexxPump 125 sends to an external controller.
The following descriptions apply for the factory setting of the parameter E (E0), see 5.1.6 E
(Error / output signal inversion).
7.2.1. Overview of output signals
The FlexxPump 125 sends the following predefined output signals to an external controller.
LCD display Designation Signal
Description
ON
Ready for
operation
high, permanent FlexxPump 125 is ready for operation
E1
Empty level
0.5 Hz,
square wave
Cartridge is empty
E2
No cartridge
low, permanent
No cartridge attached
E3
Undervoltage low, permanent
Supply voltage is too low
E4 bis E6
not occupied
E7
Overpressure low, permanent
Permissible maximum pressure reached
7.2.2. Error E1 (empty)
The error E1 signals that the cartridge is empty.
U
0 V
24 V
t
> 3 s
1 s
PIN 4
Description:
• The electronics in the FlexxPump 125 count the dosing processes according to the used
cartridge. 833 dosing processes are possible when using a 125 cm³ cartridge and 1,666
with a 250 cm³ cartridge.
• The error E1 (empty signal) can only occur after a dosing process, any follow-up dosing is
interrupted.
• The error cannot be acknowledged, the error E1 becomes error E2 after removing the
empty cartridge.
Summary of Contents for 125
Page 2: ......
Page 5: ...BA FlexxPump 125 24 V_20181121_EN de 201811 5 ...
Page 37: ...BA FlexxPump 125 24 V_20181121_EN de 201811 37 Declaration of Conformity Notes ...
Page 38: ...38 BA FlexxPump 125 24 V_20181121_EN de 201811 Declaration of Conformity Notes ...
Page 39: ...BA FlexxPump 125 24 V_20181121_EN de 201811 39 ...