Maintenance
Important!
Check all of the components at least once
a year for damage and loose connections.
Damaged or worn out parts must be exch-
anged for original spare parts immediately
by authorized experts.
Maintenance by patients
– Patients are responsible for using the
device according to the Setup and Ope-
rating Manual. Do not wash softgoods.
Maintenance between patients
– Softgoods for the device are single pati-
ent use only and cannot be washed for
reuse.
– Check the entire device for any visible
evidence of damage such as bent
components, cracked or broken covers,
frayed or damaged wires etc. If any signs
of damage are found, the device must be
repaired before use.
– Ensure that all knobs and/or levers are
usable and in place.
– Ensure that all moving components
move freely as required.
– Check all displays and electronic controls
for proper operation.
– Check all mechanical pivot and linkage
points for smooth operation and secure
mechanical connection. Make sure all
screws, nuts, bolts, rivets, pivot pins,
and other fasteners are secure.
– Gently wipe clean all exposed surfaces
with a soft cloth dampened with a mild
soap solution or alcohol. Do not use
abrasive cleaners. To dis infect, wipe
all exposed surfaces with 10% solution
of bleach and water, or other suitable
disinfectants.
– Ensure that are labels are present.
– Replace the patient softgoods kit.
– Verify that the device operates to its set
limits over several complete cycles.
– For Range of Motion (ROM) settings
verify device calibration by observing
the ROM of the device while taking a
visual reading using a geniometer at the
device’s anatomic pivot points. Compare
the ROM settings of the device with
the geniometer readings. ROM readings
should be /- 5° of the set para-
meters. If the readings do not fall within
the set parameters, the device needs
to be checked and recalibrated by a
properly trained Service Technician.
Maintenance every 6 months
– Repeat steps under “Maintenance
between patients”.
Maintenance every twelve months
– Verify electrical ground continuity where
applicable from the device frame to
ground pin of the power supply, if so
equipped, using a Safety Analyzer or
appropriate device.
– Repeat “Maintenance between patients”
procedure.
Maintenance every eighteen months
– A full inspection of the device by a
properly trained Service Technician is
recommended every 18 months.
– Repeat steps “Maintenance every twelve
months”.
– Fully inspect all internal and external
mechanical and drive components, and
repair or replace as necessary.
– Fully inspect all internal and external
electrical components (including wire
connectors and solder joints), and repair
or replace as necessary.
– Perform a complete recalibration and
subsequent check of electronic and
mechanical safety system including
Reverse-on-load function and range of
motion controls.
– Complete a final check of the device in
accordance with ORMED GmbH
MOT-301-DE_USA-GB_Rev_03-04-01-2017_DV3.indd 31
17.01.17 10:10
Summary of Contents for ARTROMOT-F
Page 2: ...MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 2 17 01 17 10 10 ...
Page 5: ...5 9 10 11 12 13 14 16 15 MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 5 17 01 17 10 10 ...
Page 21: ...21 MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 21 17 01 17 10 10 ...
Page 37: ...37 MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 37 17 01 17 10 10 ...
Page 38: ...38 Notes MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 38 17 01 17 10 10 ...
Page 39: ...39 Notes MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 39 17 01 17 10 10 ...