29
Eng
lis
h
Programming Pause
1. With power on, press the SET button.
The yellow light will appear indicating
that the controller is ready to receive
new programming. The LCD will flash all
the system icons.
2. Press the
PAUSE button
. The LCD will
only display the pause icon (hourglass)
and the two
EXT
and
FLEX hand icons
.
3. Select flexion or extension by depressing
either the
FLEX or EXT button
. Press
+ or – buttons to change the pause set-
ting from
0 to 30 seconds
. Pause may
be set independently in flexion and/or
extension.
4. When programming is complete, press
the SET button again to store the
settings in the hand-held programming
unit (yellow light will go off). Or, press
the
SET button
continue setting other
operating features.
Warm up feature
The Warm up feature starts the patient
off with
50% less range
of motion than
programmed and
gradually
works up to
the
full range of motion
over several cy-
cles (i.e. if programmed range is 0° to 120°
treatment would begin at half the range,
30° to 90°, and gradually work up to the
full range.) The device needs
10 cycles
to
the full range of motion.
m
PRECAUTION!
While Warm Up feature is on, all settings
on the hand-held programming unit are
locked.
1.
Turn off the power
from the POWER
ON/OFF switch on the hand controller.
2. Place the
Warm Up feature switch
in the ON
position and then turn the
power on from the
POWER ON/OFF
switch
again. The thermometer Warm-
up feature icon will appear in the LCD
display.
3. When ready, start the device from the
START/STOP
button.
5.2 Programming the special
functions
m
PRECAUTION!
The following are NOT for patient use.
Reverse-on-load, Lock-Out and Timer
are features to be set by the medical care
giver only. Use by the patient could result
in injury.
Reverse-on-load
The Reverse-on-load setting is program-
med by the medical care giver to meet the
treatment requirements for each patient.
This feature is hidden
from the patient for
reason and safety.
1.
Turn power off
from the POWER ON/
OFF switch.
2.
Simultaneously
press and hold the
SPEED
button and
POWER ON
switch.
The force icon will be displayed and the
flexion and extension icons will blink (-1
is lowest force setting, 5 is highest force
setting).
3. Select which range you wish to set force
in by choosing flexion or extension.
Force may be set independently in both
flexion or extension.
4. With the + or – buttons increase or
decrease force setting.
5. Press the
SET button
to program the
new force setting.
Lock-Out
The lock out feature allows the medical
Care giver to effectively lock out setting to
prevent a patient from making adjustments.
If any of the features are locked out, the
Lock-Out icon is displayed at all times.
1.
Turn power
off from POWER ON/
OFF switch.
2.
Simultaneously
press the
SET
button
and the
POWER ON
switch. The Lock-
out icon will flash.
3. Select and
press the button of the
operating features
you wish to lock
MOT-301-DE_USA-GB_Rev_03-04-01-2017_DV3.indd 29
17.01.17 10:10
Summary of Contents for ARTROMOT-F
Page 2: ...MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 2 17 01 17 10 10 ...
Page 5: ...5 9 10 11 12 13 14 16 15 MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 5 17 01 17 10 10 ...
Page 21: ...21 MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 21 17 01 17 10 10 ...
Page 37: ...37 MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 37 17 01 17 10 10 ...
Page 38: ...38 Notes MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 38 17 01 17 10 10 ...
Page 39: ...39 Notes MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 39 17 01 17 10 10 ...