![Dittmann BSM 382 Operating Manual Download Page 40](http://html1.mh-extra.com/html/dittmann/bsm-382/bsm-382_operating-manual_2510434040.webp)
Surges
according
to IEC
61000-4-5
± 1 kV
symmetri
cal
voltage
± 2 kV
asymmet
rical
voltage
Not
applicabl
e
Mains power quality
should be that of a
typical residential,
commercial or
hospital
environment.
Voltage
dips, short
interruption
s and
voltage
variations
of the
power
supply
according
to IEC
61000-4-
11
< 5 % UT
(> 95 %
voltage
dip of UT)
for ½
periods
40 % UT
(60 %
voltage
dip of UT)
for 5
periods
70 % UT
(30 %
voltage
dip of UT)
for 25
periods <
5 % UT
(> 95 %
voltage
dip of UT)
for 5 s
Not
applicabl
e
Mains power quality
should be that of a
typical residential,
commercial or
hospital
environment. If the
user of the BSM 382
requires continued
operation during
power line
interruptions, it is
recommended that
the BSM 382 be
powered from an
uninterruptible
power supply or a
battery.
Immunity
test
IEC
60601-
test
level
Confor
mity
level
Electromagnetic
environment –
guidelines
40
13.0 Information on electro-
magnetic immunity
ELECTRICAL IMMUNITY